The clinical significance of metal ion release from cobalt-chromium metal-on-metal hip joint arthroplasty

被引:92
作者
Cobb, AG
Schmalzreid, TP
机构
[1] Epsom Gen Hosp, Surrey KT18 7ET, England
[2] Orthoped Hosp, Joint Replacement Inst, Los Angeles, CA USA
关键词
metal-on-metal bearings; metal ions; total hip replacement; surface hip replacement; cobalt; chromium;
D O I
10.1243/09544119JEIM78
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Metal-on-metal (MOM) bearings offer extremely low wear and the avoidance of polyethylene but generate metallic wear particles. Although their total volume is dramatically smaller than polyethylene debris, these particles are in the nanometre size range and are many times more numerous. Metallic particles are ingested by macrophages or may be disseminated via lymphatics to the reticuloendothelial system. They corrode, and metal ions are present in the circulation and concentrated in erythrocytes. Excretion of metal ions via the kidneys seems to balance their generation in patients with MOM implants. However, highly sensitive detection methods can be used to show that levels of circulating cobalt and chromium ions are several times the normal level. These concentrations are well within the limits identified as dangerous to health in workers exposed to industrial chemicals, and also considerably lower than the levels found to cause cell toxicity in vitro. The local concentrations of particles and metal ions in the synovial tissue may occasionally exceed these limits and cause tissue necrosis. Clinical experience of lysis is rare in association with MOM bearings, as are hypersensitivity reactions and MOM bearings have had an excellent record over four decades and have a favourable benefit to risk ratio. Further reduction in risk will be achieved by improvement of materials, engineering, and accuracy of insertion.
引用
收藏
页码:385 / 398
页数:14
相关论文
共 80 条
[1]
Early clinical failure of total joint replacement in association with follicular proliferation of B-lymphocytes - A report of two cases [J].
Al-Saffar, N .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (12) :2270-2273
[2]
Metal-on-metal hybrid surface arthroplasty:: Two to six-year follow-up study [J].
Amstutz, HC ;
Beaulé, PE ;
Dorey, FJ ;
Le Duff, MJ ;
Campbell, PA ;
Gruen, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :28-39
[3]
Archibeck MJ, 2000, CLIN ORTHOP RELAT R, P12
[4]
THE MCKEE-FARRAR HIP-ARTHROPLASTY - A LONG-TERM STUDY [J].
AUGUST, AC ;
ALDAM, CH ;
PYNSENT, PB .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (04) :520-527
[5]
Cobalt [J].
Barceloux, DG .
JOURNAL OF TOXICOLOGY-CLINICAL TOXICOLOGY, 1999, 37 (02) :201-216
[6]
Chromium [J].
Barceloux, DG .
JOURNAL OF TOXICOLOGY-CLINICAL TOXICOLOGY, 1999, 37 (02) :173-194
[7]
Metal-on-metal surface arthroplasty with a cemented femoral component -: A 7-10 year follow-up study [J].
Beaulé, PE ;
Le Duff, M ;
Campbell, P ;
Dorey, FJ ;
Park, SH ;
Amstutz, HC .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :17-22
[8]
BEAULE PE, 2004, CLIN ORTHOP RELAT R, V418, P87
[9]
BEAULE PE, 2001, CLIN ORTHOP RELAT R, V386, P159
[10]
Adverse tissue reactions to wear particles form Co-alloy articulations, increased by alumina-blasting particle contamination from cementless Ti-based total hip implants -: A report of seven revisions with early failure [J].
Böhler, M ;
Kanz, F ;
Schwarz, B ;
Steffan, I ;
Walter, A ;
Plenk, H ;
Knahr, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B :128-136