A new type of sample tube for reducing convection effects in PGSE-NMR measurements of self-diffusion coefficients of liquid samples

被引:70
作者
Hayamizu, K [1 ]
Price, WS
机构
[1] Natl Inst Adv Ind Sci & Technol, AIST, Tsukuba Ctr 5, Tsukuba, Ibaraki 3058565, Japan
[2] Univ Western Sydney, Coll Sci Technol & Environm, Nanotechnol Grp, Penrith, NSW 1797, Australia
关键词
convection; diffusion; ion-ion and ion-solvent interactions; low viscosity liquid; NMR tube; PGSE-NMR;
D O I
10.1016/j.jmr.2004.01.006
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Pulsed gradient spin-echo (PGSE) NMR measurements of the self-diffusion coefficients of low viscosity liquids are greatly hampered by the effects of convection especially away from ambient temperature. Here we report on a new NMR tube designed to minimize the deleterious effects of convection. In this tube, which derives from a Shigemi symmetrical NMR tube, the sample is contained in an annulus formed from a concentric cylinder of susceptibility matched glass. The performance of this tube was demonstrated by conducting measurements on the electrochemically important LiN(So(3)CF(3))(2) (LiTFSI)-diglyme (DG) system. Calibrations were first made using DG at column heights of 2, 3, and 4-mm in the temperature range between -40 and 100degreesC. Measurements of the diffusion coefficients of the lithium, anion, and DG were then performed to probe the solvent-ion and ion-ion interactions in the DG doped with LiTFSI. Changes in the H-1, Li-7, and F-19 PGSE-NMR attenuation curves at -40degreesC provided clear evidence of interactions between the DG and lithium ion. (C) 2004 Elsevier Inc. All rights reserved.
引用
收藏
页码:328 / 333
页数:6