Self-normalizing method to measure the detective quantum efficiency of a wide range of x-ray detectors

被引:55
作者
Stierstorfer, K [1 ]
Spahn, M [1 ]
机构
[1] Siemens Med Engn, D-91301 Forchheim, Germany
关键词
detective quantum efficiency; DQE measurement method; digital radiography; noise equivalent quanta; noise power spectra;
D O I
10.1118/1.598626
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The detective quantum efficiency (DQE) is widely accepted as the most relevant parameter to characterize the image quality of medical x-ray systems. In this article we describe a solid. method to measure the DQE. The strength of the method lies in the fact that it is self-normalizing so measurements at very low spatial frequencies are not needed. Furthermore, it works on any system with a response function which is linear in the small-signal approximation. We decompose the DQE into several easily accessible quantities and discuss in detail how they can be measured. At the end we lead the interested reader through an example. Noise equivalent quanta and normalized contrast values are tabulated for standard radiation qualities. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00907-4].
引用
收藏
页码:1312 / 1319
页数:8
相关论文
共 19 条
[1]   ANALYSIS AND CORRECTION OF IMPERFECTIONS IN THE IMAGE INTENSIFIER-TV-DIGITIZER IMAGING CHAIN [J].
BOONE, JM ;
SEIBERT, JA ;
BARRETT, WA ;
BLOOD, EA .
MEDICAL PHYSICS, 1991, 18 (02) :236-242
[2]   Amorphous silicon x-ray image sensor [J].
Chabbal, J ;
Chaussat, C ;
Ducourant, T ;
Fritsch, L ;
Michailos, J ;
Spinnler, V ;
Vieux, G ;
Arques, M ;
Hahm, G ;
Hoheisel, M ;
Horbaschek, H ;
Schulz, R ;
Spahn, M .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :499-510
[3]  
Dainty JC., 1974, IMAGE SCI
[4]  
*DIN, 1983, 6872 DIN 1
[5]   DQE(F) OF 4 GENERATIONS OF COMPUTED RADIOGRAPHY ACQUISITION DEVICES [J].
DOBBINS, JT ;
ERGUN, DL ;
RUTZ, L ;
HINSHAW, DA ;
BLUME, H ;
CLARK, DC .
MEDICAL PHYSICS, 1995, 22 (10) :1581-1593
[6]   A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY [J].
FUJITA, H ;
TSAI, DY ;
ITOH, T ;
DOI, K ;
MORISHITA, J ;
UEDA, K ;
OHTSUKA, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :34-39
[7]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .1. MODULATION TRANSFER-FUNCTION [J].
GIGER, ML ;
DOI, K .
MEDICAL PHYSICS, 1984, 11 (03) :287-295
[8]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[9]  
HILLEN W, 1989, P SOC PHOTO-OPT INS, V1090, P296
[10]  
Hoeschen D, 1984, Rofo, V140, P600, DOI 10.1055/s-2008-1053034