Near-infrared spectroscopy in the routine diagnostic work-up of patients with leg ischaemia

被引:42
作者
Ubbink, DT
Koopman, B
机构
[1] Acad Med Ctr, Dept Clin Epidemiol & Biostat, NL-1100 DE Amsterdam, Netherlands
[2] Acad Med Ctr, Dept Surg, NL-1100 DE Amsterdam, Netherlands
关键词
spectroscopy; near-infrared; peripheral vascular diseases; reproducibility of results; microcirculation oximetry;
D O I
10.1016/j.ejvs.2005.10.025
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective. To examine the reproducibility and clinical applicability of near-infrared spectroscopy (NIRS) in patients with leg ischaemia. Design. Prospective comparative diagnostic study. Materials and methods. Routinely measured peripheral blood pressure and microcirculatory parameters were compared with tissue oxygen saturation (TsO2) measurements using a new NIRS device on the calf muscle and the foot. Healthy subjects (n = 20) and patients in different stages of leg ischaemia (n = 45) were investigated at rest and after provocation: a treadmill test, arterial occlusion and a change in posture. Healthy volunteers Were used as all age-matched reference population. Results. Reproducibility of the NIRS was excellent (intraclass correlation coefficient at rest was 0.91 (95% confidence interval: 80-99). Resting TsO2 (65%) in healthy controls did not differ significantly from that in patients (Fontaine 2: 66%; F3/4: 68%). After exercise, a significant reduction in TsO2 was observed only in patients: Fontaine 1: 60%; Fontaine 2: 21 %, Fontaine 3/4: 29%, The ankle/brachial index after exercise showed a good correlation (R= 0.73) with TsO2 at the end of the treadmill test. No correlation between NIRS and other micro- or macrocirculatory parameters was found. Conclusion. NIRS is a very reproducible tool to assess tissue oxygen saturation, bill is not useful for the routine work-up of patients with leg ischaemia.
引用
收藏
页码:394 / 400
页数:7
相关论文
共 28 条
[1]   PROGNOSTIC VALUE OF SYSTOLIC ANKLE AND TOE BLOOD-PRESSURE LEVELS IN OUTCOME OF DIABETIC FOOT ULCER [J].
APELQVIST, J ;
CASTENFORS, J ;
LARSSON, J ;
STENSTROM, A ;
AGARDH, CD .
DIABETES CARE, 1989, 12 (06) :373-378
[2]   The validity, reliability, reproducibility and extended utility of ankle to brachial pressure index in current vascular surgical practice [J].
Caruana, MF ;
Bradbury, AW ;
Adam, DJ .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2005, 29 (05) :443-451
[3]   NEAR-INFRARED SPECTROSCOPY IN PERIPHERAL VASCULAR-DISEASE [J].
CHEATLE, TR ;
POTTER, LA ;
COPE, M ;
DELPY, DT ;
SMITH, PDC ;
SCURR, JH .
BRITISH JOURNAL OF SURGERY, 1991, 78 (04) :405-408
[4]   DETERMINATION OF OXYGEN-CONSUMPTION IN MUSCLE DURING EXERCISE USING NEAR-INFRARED SPECTROSCOPY [J].
COLIER, WNJM ;
MEEUWSEN, IBAE ;
DEGENS, H ;
OESEBURG, B .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 :151-155
[5]   Interobserver and intraobserver reproducibility of peripheral blood and oxygen pressure measurements in the assessment of lower extremity arterial disease [J].
de Graaff, J ;
Ubbink, DT ;
Legemate, DA ;
de Haan, RJ ;
Jacobs, MJHM .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (05) :1033-1040
[6]   The usefulness of a laser Doppler in the measurement of the blood pressures [J].
de Graaff, JC ;
Ubbink, DT ;
Legemate, DA ;
de Haan, RJ ;
Jacobs, MJHM .
JOURNAL OF VASCULAR SURGERY, 2000, 32 (06) :1172-1179
[7]   The severity of muscle ischemia during intermittent claudication [J].
Egun, A ;
Farooq, V ;
Torella, F ;
Cowley, R ;
Thorniley, MS ;
McCollum, CN .
JOURNAL OF VASCULAR SURGERY, 2002, 36 (01) :89-93
[8]   EQUIVALENCE OF WEIGHTED KAPPA AND INTRACLASS CORRELATION COEFFICIENT AS MEASURES OF RELIABILITY [J].
FLEISS, JL ;
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1973, 33 (03) :613-619
[9]   Utility of near-infrared spectroscopy in the diagnosis of lower extremity compartment syndrome [J].
Giannotti, G ;
Cohn, SM ;
Brown, M ;
Varela, JE ;
McKenney, MG ;
Wiseberg, JA .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2000, 48 (03) :396-399
[10]   NONINVASIVE, INFRARED MONITORING OF CEREBRAL AND MYOCARDIAL OXYGEN SUFFICIENCY AND CIRCULATORY PARAMETERS [J].
JOBSIS, FF .
SCIENCE, 1977, 198 (4323) :1264-1267