Clinical Assessment of Metal Artifact Reduction Methods in Dual-Energy CT Examinations of Instrumented Spines

被引:35
作者
Long, Zaiyang [1 ]
DeLone, David R. [1 ]
Kotsenas, Amy L. [1 ]
Lehman, Vance T. [1 ]
Nagelschneider, Alex A. [1 ]
Michalak, Gregory J. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 1st St SW, Rochester, MN 55905 USA
关键词
dual-energy CT; iMAR; metal artifact; spine; virtual monochromatic imaging; ITERATIVE RECONSTRUCTION; ORTHOPEDIC IMPLANTS; CLIP ARTIFACTS; FUSION; ALGORITHM; HARDWARE;
D O I
10.2214/AJR.18.19757
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
OBJECTIVE. The purpose of this study was to evaluate the performance of three metal artifact reduction methods in dual-energy CT (DECT) examinations of instrumented spines. MATERIALS AND METHODS. Twenty patients with instrumented spines who underwent spine DECT were retrospectively identified. All scans were obtained on a dual-source 128-MDCT scanner. In addition to the original DE mixed images, DECT images were reconstructed using an iterative metal artifact reconstruction algorithm (DE iMAR), virtual monochromatic imaging (VMI) algorithm (DE Mono+), and a combination of the two algorithms DE iMAR and DE Mono+, which we refer to here as DE iMAR Mono+. The four image series were anonymized and randomized for a reader study. Four experienced neuroradiologists rated the images in terms of artifact scores of four anatomic regions and overall image quality scores in both bone and soft-tissue display window settings. In addition, a quantitative analysis was performed to assess the performance of the three metal artifact reduction methods. RESULTS. There were statistically significant differences in the artifact scores and overall image quality scores among the four methods (both, p < 0.001). DE iMAR Mono+ showed the best artifact scores and quality scores (all, p < 0.001). The intraclass correlation coefficient for the overall image quality score was 0.779 using the bone display window and 0.892 using the soft-tissue display window (both, p < 0.001). In addition, DE iMAR Mono+ reduced the artifacts by the greatest amount in the quantitative analysis. CONCLUSION. The method that used DE iMAR Mono+ showed the best performance of spine metal artifact reduction using DECT data. These results may be specific to this CT vendor and implant type.
引用
收藏
页码:395 / 401
页数:7
相关论文
共 33 条
[1]
Metal artifact reduction strategies for improved attenuation correction in hybrid PET/CT imaging [J].
Abdoli, Mehrsima ;
Dierckx, Rudi A. J. O. ;
Zaidi, Habib .
MEDICAL PHYSICS, 2012, 39 (06) :3343-3360
[2]
Postoperative magnetic resonance imaging artifact with cobalt-chromium versus titanium spinal instrumentation [J].
Ahmad, Faiz U. ;
Sidani, Charif ;
Fourzali, Roberto ;
Wang, Michael Y. .
JOURNAL OF NEUROSURGERY-SPINE, 2013, 19 (05) :629-636
[3]
ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[4]
Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors [J].
Andersson, K. M. ;
Nowik, P. ;
Persliden, J. ;
Thunberg, P. ;
Norrman, E. .
BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1052)
[5]
Comparison and Combination of Dual-Energy- and Iterative-Based Metal Artefact Reduction on Hip Prosthesis and Dental Implants [J].
Bongers, Malte N. ;
Schabel, Christoph ;
Thomas, Christoph ;
Raupach, Rainer ;
Notohamiprodjo, Mike ;
Nikolaou, Konstantin ;
Bamberg, Fabian .
PLOS ONE, 2015, 10 (11)
[6]
Dual-Energy Multidetector CT: How Does It Work, What Can It Tell Us, and When Can We Use It in Abdominopelvic Imaging? [J].
Coursey, Courtney A. ;
Nelson, Rendon C. ;
Boll, Daniel T. ;
Paulson, Erik K. ;
Ho, Lisa M. ;
Neville, Amy M. ;
Marin, Daniele ;
Gupta, Rajan T. ;
Schindera, Sebastian T. .
RADIOGRAPHICS, 2010, 30 (04) :1037-1055
[7]
United States trends in lumbar fusion surgery for degenerative conditions [J].
Deyo, RA ;
Gray, DT ;
Kreuter, W ;
Mirza, S ;
Martin, BI .
SPINE, 2005, 30 (12) :1441-1445
[8]
Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT [J].
Dong, Yue ;
Shi, Ai Jun ;
Wu, Jian Lin ;
Wang, Ru Xin ;
Sun, Li Fei ;
Liu, Ai Lian ;
Liu, Yi Jun .
EUROPEAN SPINE JOURNAL, 2016, 25 (06) :1754-1763
[9]
Multichannel CT: Evaluating the spine in postoperative patients with orthopedic hardware [J].
Douglas-Akinwande, Annette C. ;
Buckwalter, Kenneth A. ;
Rydberg, Jonas ;
Rankin, James L. ;
Choplin, Robert H. .
RADIOGRAPHICS, 2006, 26 :S97-U107
[10]
Gjesteby L, 2017, P SPIE MED IMAGING, P10132