Comparison of two plethysmography systems in assessment of forearm blood flow

被引:19
作者
Leslie, SJ
Attinà, T
Hultsch, E
Bolscher, L
Grossman, M
Denvir, MA
Webb, DJ [1 ]
机构
[1] Univ Edinburgh, Western Gen Hosp, Clin Pharmacol Unit, Edinburgh EH4 2XU, Midlothian, Scotland
[2] Univ Edinburgh, Western Gen Hosp, Res Ctr, Edinburgh EH4 2XU, Midlothian, Scotland
[3] Univ Dresden, Dept Pharmacol, D-01307 Dresden, Germany
关键词
validation; Stroop's colored word conflict test; lower body negative pressure;
D O I
10.1152/japplphysiol.00567.2002
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Venous occlusion plethysmography is widely used to assess forearm blood flow (FBF). We compared the established Hokanson system (HEC4) with a newly developed Filtrass 2001 system (F2001). The HEC4 uses mercury-in-Silastic strain gauges, whereas F2001 detects volume changes with a nonmercury linear displacement device. The aim of this study was to evaluate the new F2001 against the HEC4 in terms of repeatability and systematic bias. Ten subjects were studied on 4 separate days in random order using either the HEC4 on both arms, the F2001 on both arms, the HEC4 on the right arm with the F2001 on the left, or the F2001 on the right arm and the HEC4 on the left. Stroop's colored word conflict test and postocclusive hyperemia were used to increase FBF, and lower body negative pressure was used to lower FBF. Stroop's colored word conflict test and lower body negative pressure increased (24.6+/-1.5%, n=240, P<0.0001) and decreased (18.7 +/- 0.8%, n=240, P<0.0001) FBF, respectively. Postocclusive hyperemia after occlusion times of 5, 8, and 13 min substantially increased FBF by 390+/-86, 756+/-217, and 851+/-132%, respectively. Repeatability was not different between the devices (0.10+/-2.37 vs. -0.47+/-1.92 l/min, n=125, P>0.05), and there was no systematic bias. The F2001 is a newly developed plethysmography system that does not utilize mercury and is suitable for assessing changes of FBF in physiological studies.
引用
收藏
页码:1794 / 1799
页数:6
相关论文
共 25 条
[1]   LOCAL INHIBITION OF CONVERTING ENZYME AND VASCULAR-RESPONSES TO ANGIOTENSIN AND BRADYKININ IN THE HUMAN FOREARM [J].
BENJAMIN, N ;
COCKCROFT, JR ;
COLLIER, JG ;
DOLLERY, CT ;
RITTER, JM ;
WEBB, DJ .
JOURNAL OF PHYSIOLOGY-LONDON, 1989, 412 :543-555
[2]   MEASURING FOREARM BLOOD-FLOW AND INTERPRETING THE RESPONSES TO DRUGS AND MEDIATORS [J].
BENJAMIN, N ;
CALVER, A ;
COLLIER, J ;
ROBINSON, B ;
VALLANCE, P ;
WEBB, D .
HYPERTENSION, 1995, 25 (05) :918-923
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
BRODIE TG, 2005, P PHYS SOC LOND PHYS
[5]  
CHANG P, 1987, J CARDIOVASC PHARM, V10, P123
[6]   Description and validation of a novel liquid metal-free device for venous congestion plethysmography [J].
Christ, F ;
Bauer, A ;
Brügger, D ;
Niklas, M ;
Gartside, IB ;
Gamble, J .
JOURNAL OF APPLIED PHYSIOLOGY, 2000, 89 (04) :1577-1583
[7]   Reproducibility of resting peripheral blood flow using strain gauge plethysmography [J].
Fehling, PC ;
Arciero, PJ ;
MacPherson, CJ ;
Smith, DL .
INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1999, 20 (08) :555-559
[8]   CARDIOVASCULAR AND SYMPATHOADRENAL RESPONSES TO MENTAL STRESS - INFLUENCE OF BETA-BLOCKADE [J].
FREYSCHUSS, U ;
HJEMDAHL, P ;
JUHLINDANNFELT, A ;
LINDE, B .
AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (06) :H1443-H1451
[9]   REACTIONS OF THE BLOOD VESSELS OF THE HUMAN FOREARM TO INCREASES IN TRANSMURAL PRESSURE [J].
GREENFIELD, ADM ;
PATTERSON, GC .
JOURNAL OF PHYSIOLOGY-LONDON, 1954, 125 (03) :508-524
[10]   ELECTRICALLY CALIBRATED PLETHYSMOGRAPH FOR DIRECT MEASUREMENT OF LIMB BLOOD-FLOW [J].
HOKANSON, DE ;
SUMNER, DS ;
STRANDNESS, DE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1975, BM22 (01) :25-29