The AO/ASIF-proximal femoral nail (PFN): a new device for the treatment of unstable proximal femoral fractures

被引:168
作者
Simmermacher, RKJ [1 ]
Bosch, AM [1 ]
Van der Werken, C [1 ]
机构
[1] Univ Utrecht Hosp, Dept Surg, NL-3508 GA Utrecht, Netherlands
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 1999年 / 30卷 / 05期
关键词
D O I
10.1016/S0020-1383(99)00091-1
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Currently available intramedullary devices for the treatment of unstable pertrochanteric femoral fractures are technically demanding and show a considerable failure rate due to their massiveness and the persistent rotational instability of the head-neck fragment. The Arbeitsgemeinschaft fur Osteosynthesefragen (AO/ASIF) therefore designed a new intramedullary device, the proximal femoral nail (PFN), which during a handling study was tested in 4 European clinics. Within one year 191 proximal femoral fractures were treated with the PFN. After a follow up period of at least 4 months technical failures (poor reduction, malrotation or wrong choice of screws) were seen in just 4.6% of the cases. One patient showed a cut-out of the implant but there were no mechanical failures (bending, breaking of the implant) or ipsilateral fractures of the femoral shaft at the tip of the implant. It is therefore concluded that the results of this new implant compare favourably to the currently available implants for the treatment of the unstable pertrochanteric femoral fracture. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:327 / 332
页数:6
相关论文
共 24 条
[1]   Complications and technical problems with the gamma nail [J].
Albareda, J ;
Laderiga, A ;
Palanca, D ;
Paniagua, L ;
Seral, F .
INTERNATIONAL ORTHOPAEDICS, 1996, 20 (01) :47-50
[2]   GAMMA-NAIL VS COMPRESSION SCREW FOR TROCHANTERIC FEMORAL FRACTURES - 15 REOPERATIONS IN A PROSPECTIVE, RANDOMIZED STUDY OF 378 PATIENTS [J].
AUNE, AK ;
EKELAND, A ;
ODEGAARD, B ;
GROGAARD, B ;
ALHO, A .
ACTA ORTHOPAEDICA SCANDINAVICA, 1994, 65 (02) :127-130
[3]   FIXATION OF INTERTROCHANTERIC FRACTURES OF THE FEMUR - A RANDOMIZED PROSPECTIVE COMPARISON OF THE GAMMA-NAIL AND THE DYNAMIC HIP SCREW [J].
BRIDLE, SH ;
PATEL, AD ;
BIRCHER, M ;
CALVERT, PT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (02) :330-334
[4]   COMPARISON OF DYNAMIC HIP SCREW AND GAMMA-NAIL - A PROSPECTIVE, RANDOMIZED, CONTROLLED TRIAL [J].
BUTT, MS ;
KRIKLER, SJ ;
NAFIE, S ;
ALI, MS .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1995, 26 (09) :615-618
[5]   TROCHANTERIC FRACTURES OF THE FEMUR - A RANDOMIZED PROSPECTIVE TRIAL COMPARING THE JEWETT NAIL-PLATE WITH THE DYNAMIC HIP SCREW [J].
ESSER, MP ;
KASSAB, JY ;
JONES, DHA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (04) :557-560
[6]  
EULER E, 1996, HEFTE UNFALLCHIR, V262, P2
[7]  
FRIEDL W, 1994, CHIRURG, V65, P953
[8]  
FRIEDL W, 1996, HEFTE UNFALLCHIR, V262, P4
[9]  
GUYER P, 1991, TRAUMATOL, V21, P242
[10]   THE GAMMA-NAIL FOR PERITROCHANTERIC FRACTURES [J].
HALDER, SC .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :340-344