Design optimisation and experimental evaluation of dorsal double plating fixation for distal radius fracture

被引:14
作者
Chen, Alvin Chao-Yu [1 ,2 ]
Lin, Yu-Hao [3 ]
Kuo, Hsien-Nan [4 ]
Yu, Tsung-Chih [4 ]
Sun, Ming-Tsung [3 ]
Lin, Chun-Li [5 ]
机构
[1] Chang Gung Mem Hosp, Dept Orthopaed Surg, Tao Yuan 333, Taiwan
[2] Univ Coll Med, Tao Yuan 333, Taiwan
[3] Chang Gung Univ, Dept Mech Engn, Tao Yuan 333, Taiwan
[4] Met Ind Res & Dev Ctr, Med Devices Dev Sect, Lujhu Township 82151, Kaohsiung Count, Taiwan
[5] Natl Yang Ming Univ, Dept Biomed Engn, Taipei 112, Taiwan
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2013年 / 44卷 / 04期
关键词
Dorsal double plating; Finite element; Taguchi; Optimisation; Fatigue; INTERNAL-FIXATION; EXTERNAL FIXATION; COLLES FRACTURES; T-PLATES; VOLAR; LOCKING; MODEL; BIOMECHANICS; STABILITY; FORCES;
D O I
10.1016/j.injury.2012.09.022
中图分类号
R4 [临床医学];
学科分类号
100218 [急诊医学];
摘要
This study determines the relative effects of changes in osteoporosis condition, plate/screw design factors (plate angle/length/width/thickness and screw diameter) and fixation methods (screw number and screw length) on the biomechanical response of dorsal double plating (DDP) fixation at a distal radius fracture to determine the optimal design and evaluate its biomechanical strength using the dynamic fatigue test. Eighteen CAD and finite element (FE) models corresponding to a Taguchi L-18 array were constructed to perform numerical simulations to simulate the mechanical responses of a DDP fixed in a simply distal radius fracture bone. The Taguchi method was employed to determine the significance of each design factor in controlling bone/plate/screw stress and distal fragment displacement under axial (100 N), bending (1 N m) and torsion (1 N m) loads. Simulation results indicated that the order rank to determine the mechanical response was the plate thickness, plate width, screw diameter, and number of screws. Dorsal intermediate (L) plate with 60 mm length, 1.8 mm thickness, 6.0 mm width and 2.8 mm diameter, 20 mm length dual-thread locking screw can be found for optimisation. The DDP, including an L plate with 0 degrees, 30 degrees and 60 degrees angles and a straight I plate, were made with Ti6Al4V to fix onto the sawbones with three corresponding radius fractures to perform the dynamic testing. The specimens were oscillated with loads between 10 N and 150 N at 5 Hz for 20,000 cycles. The average stiffness in 20,000 test cycles was 425.7 N/mm, 461.1 N/mm and 532.1 N/mm for the 0 degrees, 30 degrees and 60 degrees constructs, respectively. No difference in stiffness was found in the same angled constructs throughout the 20,000 cycles of testing (p > 0.05). Lack of gross construct failures during cyclic testing and reasonable stiffness corroborated that our new constructs tested to date seem stable enough to support restricted postoperative loads. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:527 / 534
页数:8
相关论文
共 34 条
[11]
EXTERNAL FIXATION OR PLASTER FOR SEVERELY DISPLACED COMMINUTED COLLES FRACTURES - A PROSPECTIVE-STUDY OF ANATOMICAL AND FUNCTIONAL RESULTS [J].
HOWARD, PW ;
STEWART, HD ;
HIND, RE ;
BURKE, FD .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1989, 71 (01) :68-73
[12]
Fractures of the distal radius treated by internal fixation and early function - A prospective study of 73 consecutive patients [J].
Jakob, M ;
Rikli, DA ;
Regazzoni, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (03) :340-344
[13]
FRACTURES OF THE DISTAL END OF THE RADIUS [J].
JUPITER, JB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (03) :461-469
[14]
Does a volar locking plate provide equivalent stability as a dorsal nonlocking plate in a dorsally comminuted distal radius fracture? a biomechanical study [J].
Kandemir, Utku ;
Matityahu, Amir ;
Desai, Rohan ;
Puttlitz, Christian .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (09) :605-610
[15]
Lin YH, J MED BIOL IN PRESS
[16]
A biomechanical comparison of a dorsal 3.5-mm T-plate and a volar fixed-angle plate in a model of dorsally unstable distal radius fractures [J].
Liporace, FA ;
Gupta, S ;
Jeong, GK ;
Stracher, M ;
Kummer, F ;
Egol, KA ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2005, 19 (03) :187-191
[17]
A biomechanical comparison of two volar locked plates in a dorsally unstable distal radius fracture model [J].
Liporace, Frank Anthony ;
Kubiak, Erik N. ;
Jeong, Gerard K. ;
Iesaka, Kazuho ;
Egol, Kenneth A. ;
Koval, Kenneth J. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 61 (03) :668-672
[18]
Volar plating of AO C3 distal radius fractures: Biomechanical evaluation of locking screw and locking smooth peg configurations [J].
Martineau, P. A. ;
Waitayawinyu, T. ;
Malone, K. J. ;
Hanel, D. P. ;
Trumble, T. E. .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2008, 33A (06) :827-834
[19]
Comparison of different distal radius dorsal and volar fracture fixation plates: A biomechanical study [J].
Osada, D ;
Viegas, SF ;
Shah, MA ;
Morris, RP ;
Patterson, RM .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2003, 28A (01) :94-104
[20]
Biomechanics in uniaxial compression of three distal radius volar plates [J].
Osada, D ;
Fujita, S ;
Tamai, K ;
Iwamoto, A ;
Tomizawa, K ;
Saotome, K .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2004, 29A (03) :446-451