Wear of the polyethylene liner-metallic shell interface in modular acetabular components - An in vitro analysis

被引:32
作者
Lieberman, JR
Kay, RM
Hamlet, WP
Park, SH
Kabo, JM
机构
[1] Department of Orthopaedic Surgery, UCLA Medical Center, Los Angeles
[2] Veterans Administration Hospital, West Los Angeles
[3] Biomechanics Laboratory, J. Vernon Luck Sr. MD Orthopaedic Research Center, Orthopaedic Hospital, Los Angeles, CA
[4] Department of Orthopaedic Surgery, CHS 76-134, UCLA Medical Center, Los Angeles, CA 90024
关键词
total hip arthroplasty; polyethylene; wear; modularity; acetabular components;
D O I
10.1016/S0883-5403(96)80116-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of this study was to determine the effect of compression cycles and wear patterns on the polyethylene liner-metallic shell interface in modular acetabular components. Articular frictional torque was also measured. Modular acetabular components from five manufacturers were tested. The polyethylene liners were sputter coated with gold on the convex surface to enhance the visualization of wear and deformation patterns. Each component was cycled for 10 million cycles in a hydraulic fatigue testing machine. Frictional torque was measured prior to the start of the cyclic loading and in increments of 2 million cycles. Frictional torque was significantly lower in the design with an enhanced polyethylene liner. Abrasion of the gold from the convex surface of polyethylene liners varied from 2 to 23% between designs. Extrusion of the polyethylene into the screw holes in the metallic shells was universal. Three modes of damage (burnishing, punch-out, and gouging) were identified on the convex surface of the polyethylene liner. Abrasion of the gold from the convex surface of the polyethylene varied greatly between designs. This is indicative of relative motion between the polyethylene liner and the metallic shell. This motion must be minimized to limit the generation of wear debris from the convex surface of the polyethylene. Several aspects of modular acetabular component design could be implemented to potentially reduce wear, including limiting the number of holes available for screw placement, smoothing out the edges of the screw holes to void punch-out, and avoiding sup plemental fixation of the liner unless it is essential to prevent motion between the liner and the metallic shell.
引用
收藏
页码:602 / 608
页数:7
相关论文
共 12 条
[1]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[2]   DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES [J].
BRIEN, WW ;
SALVATI, EA ;
WRIGHT, TM ;
NELSON, CL ;
HUNGERFORD, DS ;
GILLIAM, DL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) :1548-1550
[3]  
Bueche M J, 1989, J Arthroplasty, V4, P39, DOI 10.1016/S0883-5403(89)80052-X
[4]  
COLLIER JP, 1992, CLIN ORTHOP RELAT R, V285, P129
[5]  
COLLIER JP, 1990, CLIN ORTHOP RELAT R, V261, P107
[6]  
HUANG DD, 1992, 38 ANN M ORTH RES SO
[7]   POLYETHYLENE AND METAL DEBRIS GENERATED BY NON-ARTICULATING SURFACES OF MODULAR ACETABULAR COMPONENTS [J].
HUK, OL ;
MANJULA, B ;
BETTS, F ;
RIMNAC, CM ;
LIEBERMAN, JR ;
HUO, MH ;
SALVATI, EA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04) :568-574
[8]   DISASSEMBLY OF A MODULAR ACETABULAR COMPONENT OF A TOTAL HIP-REPLACEMENT ARTHROPLASTY - A CASE-REPORT [J].
KITZIGER, KJ ;
DELEE, JC ;
EVANS, JA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :621-623
[9]   FRICTIONAL TORQUE IN SURFACE AND CONVENTIONAL HIP-REPLACEMENT [J].
MA, SM ;
KABO, JM ;
AMSTUTZ, HC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1983, 65 (03) :366-370
[10]  
MCKELLOP H, 1992, 38 ANN M ORTH RES SO