High resolution GSO block detectors using PMT-Quadrant-Sharing design for small animal PET

被引:8
作者
Kim, S [1 ]
Wong, WH [1 ]
Xie, SP [1 ]
Ramirez, R [1 ]
Liu, YQ [1 ]
Li, HD [1 ]
Xing, T [1 ]
Wang, Y [1 ]
Uribe, J [1 ]
Baghaei, H [1 ]
Zhang, YX [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Houston, TX 77030 USA
基金
美国国家卫生研究院;
关键词
GSO block detector; PQS technology; small animal PET;
D O I
10.1109/TNS.2005.862969
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
GSO position-sensitive block detectors have been developed using the PMT-Quadrant-Sharing (PQS) technology for animal PET imaging. Two prototype block detectors with dimensions of 1.51 mm x 1.51 mm x 10 mm (12 x 12 array) and 1.66 mm x 1.66 mm x 10 mm (11 x 11 array) were built using 19 mm round PMTs. The Enhanced Specular Reflector mirror-film (98% reflectance, 0.065 mm thickness, 3M Co.) was used to control the light collection and distribution in the PQS design. The detector pitches are 1.59 (12 x 12) and 1.74 mm (11 x 11) and the crystal-packing fractions 95.0 and 95.4%, respectively. List-mode measurements with Cs-137 sources were carried out to investigate the crystal decoding and analyzed to extract the individual crystal spectra. All the crystals in both the detectors were clearly identified and well isolated. The light-collection efficiency is lowest in the central crystals and highest in the crystals of the four corners, with a ratio of 0.76 for the 11 x 11 array. The overall energy resolution of the 11 x 11 array is 13.5% (15.4% at 511 keV) with a standard deviation of 1.2%, which indicates that the individual energy resolutions of the detector are high and uniformly distributed. From this study, we achieved a high decodable crystals/PMT ratio of 144:1 for the 19 mm PQS GSO detector technology with crystals 10 rum deep. It is believed that these detectors would be useful for high resolution animal PET imaging.
引用
收藏
页码:40 / 43
页数:4
相关论文
共 20 条
[1]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[2]   Optical fiber readout of scintillator arrays using a multi-channel PMT: A high resolution PET detector for animal imaging [J].
Cherry, SR ;
Shao, YP ;
Siegel, S ;
Silverman, RW ;
Mumcuoglu, E ;
Meadors, K ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1932-1937
[3]   CERIUM DOPED GSO SCINTILLATORS AND ITS APPLICATION TO POSITION-SENSITIVE DETECTORS [J].
ISHIBASHI, H ;
SHIMIZU, K ;
SUSA, K .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :170-172
[4]  
Laforest R, 2004, IEEE NUCL SCI CONF R, P2965
[5]  
LI H, 2003, P IEEE NUCL SCI S C, V4, P2267
[6]   FEASIBILITY STUDY OF AN AVALANCHE PHOTODIODE READOUT FOR A HIGH-RESOLUTION PET WITH NSEC TIME RESOLUTION [J].
SCHMELZ, C ;
BRADBURY, SM ;
HOLL, I ;
LORENZ, E ;
RENKER, D ;
ZIEGLER, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1080-1084
[7]   Optimizing the performance of a PET detector using discrete GSO crystals on a continuous lightguide [J].
Surti, S ;
Karp, JS ;
Freifelder, R ;
Liu, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) :1030-1036
[8]   Design evaluation of A-PET: A high sensitivity animal PET camera [J].
Surti, S ;
Karp, JS ;
Perkins, AE ;
Freifelder, R ;
Muehllehner, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) :1357-1363
[9]   CERIUM-ACTIVATED GD2SIO5 SINGLE-CRYSTAL SCINTILLATOR [J].
TAKAGI, K ;
FUKAZAWA, T .
APPLIED PHYSICS LETTERS, 1983, 42 (01) :43-45
[10]   Effect of the rotational orientation of circular photomultipliers in a PET camera block detector design [J].
Uribe, J ;
Wong, WH ;
Hu, GJ ;
Hicks, K ;
Wang, JM ;
Baghaei, H ;
Zhang, N ;
Li, HD ;
Yokoyama, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1266-1270