PET/MRI system design

被引:106
作者
Delso, Gaspar [1 ]
Ziegler, Sibylle [1 ]
机构
[1] Tech Univ Munich, Nukl Med Klin & Poliklin, Klinikum Rechts Isar, D-81675 Munich, Germany
关键词
Positron emission tomography; Magnetic resonance imaging; Instrumentation; Equipment design; POSITRON-EMISSION-TOMOGRAPHY; SIMULTANEOUS PET; SIMULTANEOUS ACQUISITION; DETECTOR; SCANNER; MRI; PERFORMANCE; READOUT; IMAGES;
D O I
10.1007/s00259-008-1008-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
The combination of clinical MRI and PET systems has received increased attention in recent years. In contrast to currently used PET/CT systems, PET/MRI offers not only improved soft-tissue contrast and reduced levels of ionizing radiation, but also a wealth of MRI-specific information such as functional, spectroscopic and diffusion tensor imaging. Combining PET and MRI, however, has proven to be very challenging, due to the detrimental cross-talk effects between the two systems. Significant progress has been made in the recent years to overcome these difficulties, with several groups reporting PET/MRI prototypes for animal imaging and a clinical insert for neurological applications being demonstrated at the 2007 Annual Meeting of the Society of Nuclear Medicine. In this paper we review different architectures for clinical PET/MRI systems, and their possibilities, limitations and technological obstacles.
引用
收藏
页码:86 / 92
页数:7
相关论文
共 35 条
[1]
Beyer T, 2000, J NUCL MED, V41, P1369
[2]
Simultaneous in vivo positron emission tomography and magnetic resonance imaging [J].
Catana, Ciprian ;
Procissi, Daniel ;
Wu, Yibao ;
Judenhofer, Martin S. ;
Qi, Jinyi ;
Pichler, Bernd J. ;
Jacobs, Russell E. ;
Cherry, Simon R. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2008, 105 (10) :3705-3710
[3]
Catana C, 2006, J NUCL MED, V47, P1968
[4]
A hybrid PET-MRI: An integrated molecular-genetic imaging system with HRRT-PET and 7.0-T MRI [J].
Cho, Zang-Hee ;
Son, Young-Don ;
Kim, Hang-Keun ;
Kim, Kyoung-Nam ;
Oh, Se-Hong ;
Han, Jae-Yong ;
Hong, In-Ki ;
Kim, Young-Bo .
INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 2007, 17 (04) :252-265
[5]
POSITRON EMISSION TOMOGRAPHY WITHIN A MAGNETIC-FIELD USING PHOTOMULTIPLIER TUBES AND LIGHTGUIDES [J].
CHRISTENSEN, NL ;
HAMMER, BE ;
HEIL, BG ;
FETTERLY, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) :691-697
[6]
EFFECTS OF MAGNETIC-FIELDS ON THE LIGHT YIELD OF SCINTILLATORS [J].
CUMALAT, JP ;
CHEUNG, HWK ;
HASSED, J ;
SMITH, BD ;
BROSS, AD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 293 (03) :606-614
[7]
Characteristics of neurotransmitter competition studies using constant infusion of tracer [J].
Endres, CJ ;
Carson, RE .
QUANTITATIVE FUNCTIONAL BRAIN IMAGING WITH POSITRON EMISSION TOMOGRAPHY, 1998, :435-439
[8]
*EUR COMM, 2001, REF GUID IM RAD PROT
[9]
APD-based PET detector for simultaneous PET/MR imaging [J].
Grazioso, Ronald ;
Zhang, Nan ;
Corbeil, James ;
Schmand, Matthias ;
Ladebeck, Ralf ;
Vester, Markus ;
Schnur, Guenter ;
Renz, Wolfgang ;
Fischer, Hubertus .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02) :301-305
[10]
HANDLER W, 2007, J NUCL MED S2, V48, pP89