The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: experimental and finite-element analysis

被引:77
作者
Ferguson, SJ [1 ]
Visser, JMA [1 ]
Polikeit, A [1 ]
机构
[1] Univ Bern, Inst Surg Technol & Biomech, MEM Res Ctr, CH-3014 Bern, Switzerland
关键词
PEEK; material properties; creep; cages; fusion;
D O I
10.1007/s00586-005-0915-5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Polyetheretherketone (PEEK) is a novel polymer with potential advantages for its use in demanding orthopaedic applications (e.g. intervertebral cages). However, the influence of a physiological environment on the mechanical stability of PEEK has not been reported. Furthermore, the suitability of the polymer for use in highly stressed spinal implants such as intervertebral cages has not been investigated. Therefore, a combined experimental and analytical study was performed to address these open questions. A quasi-static mechanical compression test was performed to compare the initial mechanical properties of PEEK-OPTIMA polymer in a dry, room-temperature and in an aqueous, 37 degrees C environment (n=10 per group). The creep behaviour of cylindrical PEEK polymer specimens (n=6) was measured in a simulated physiological environment at an applied stress level of 10 MPa for a loading duration of 2000 hours (12 weeks). To compare the biomechanical performance of different intervertebral cage types made from PEEK and titanium under complex loading conditions, a three-dimensional finite element model of a functional spinal unit was created. The elastic modulus of PEEK polymer specimens in a physiological environment was 1.8% lower than that of specimens tested at dry, room temperature conditions (P < 0.001). The results from the creep test showed an average creep strain of less than 0.1% after 2000 hours of loading. The finite element analysis demonstrated high strain and stress concentrations at the bone/implant interface, emphasizing the importance of cage geometry for load distribution. The stress and strain maxima in the implants were well below the material strength limits of PEEK. In summary, the experimental results verified the mechanical stability of the PEEK-OPTIMA polymer in a simulated physiological environment, and over extended loading periods. Finite element analysis supported the use of PEEK-OPTIMA for load-bearing intervertebral implants.
引用
收藏
页码:149 / 156
页数:8
相关论文
共 39 条
[1]
Tensile properties, tension-tension fatigue and biological response of polyetheretherketone-hydroxyapatite composites for load-bearing orthopedic implants [J].
Abu Bakar, MS ;
Cheng, MHW ;
Tang, SM ;
Yu, SC ;
Liao, K ;
Tan, CT ;
Khor, KA ;
Cheang, P .
BIOMATERIALS, 2003, 24 (13) :2245-2250
[2]
Akay M, 1995, Proc Inst Mech Eng H, V209, P93, DOI 10.1243/PIME_PROC_1995_209_325_02
[3]
Akay M, 1996, J BIOMED MATER RES, V31, P167
[4]
ALBERT K, 1994, BIO-MED MATER ENG, V4, P199
[5]
*ASTM, 2004, D69502 ASTM
[6]
*ASTM, 2004, D299001 ASTM
[7]
Quantitative radiographic analysis of fiber reinforced polymer composites [J].
Baidya, KP ;
Ramakrishna, S ;
Rahman, M ;
Ritchie, A .
JOURNAL OF BIOMATERIALS APPLICATIONS, 2001, 15 (03) :279-289
[8]
CHARACTERIZATION OF SHORT-FIBER REINFORCED THERMOPLASTICS FOR FRACTURE FIXATION DEVICES [J].
BROWN, SA ;
HASTINGS, RS ;
MASON, JJ ;
MOET, A .
BIOMATERIALS, 1990, 11 (08) :541-547
[9]
Measurement at low strain rates of the elastic properties of dental polymeric materials [J].
Chabrier, F ;
Lloyd, CH ;
Scrimgeour, SN .
DENTAL MATERIALS, 1999, 15 (01) :33-38
[10]
Preliminary experience using a polyetheretherketone (PEEK) cage in the treatment of cervical disc disease [J].
Cho, DY ;
Liau, WR ;
Lee, WY ;
Liu, JT ;
Chiu, CL ;
Sheu, PC .
NEUROSURGERY, 2002, 51 (06) :1343-1349