Packed-column supercritical fluid chromatography for the purity analysis of clevidipine, a new dihydropyridine drug

被引:20
作者
Gyllenhaal, O [1 ]
Karlsson, A [1 ]
Vessman, J [1 ]
机构
[1] AstraZeneca R&D Molndal, Dept Analyt Chem, S-43183 Molndal, Sweden
关键词
purity analysis; clevidipine;
D O I
10.1016/S0021-9673(99)00921-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this paper we describe a packed column supercritical fluid chromatography method that can be used for the analysis of a new dihydropyridine substance. The method is based on methanol-modified carbon dioxide as the mobile phase and Hypersil bare silica as column support at a column temperature of 50 degrees C and 150 bar as back pressure. Using an adjusted methanol gradient the most likely by-products can be separated and detected (240 nm) within 13 min. Occasionally the column needed treatment with 4 mM citric acid in the methanol modifier in order to give a narrow peak of an acidic analogue. The present method can detect analogues at the 0.1% (w/w) level. The precision at this level for one of the analogues was 5.9% RSD. This method shows a higher selectivity than a corresponding reversed-phase liquid chromatographic method. (C) 1999 Elsevier Science B.V. All rights reserved.