Characterising differences between measurement and calibration wafer in probe-tip calibrations

被引:16
作者
Carchon, G
Nauwelaers, B
De Raedt, W
Schreurs, D
Vandenberghe, S
机构
[1] Katholieke Univ Leuven, TELEMIC, ESAT, B-3001 Heverlee, Belgium
[2] IMEC, Div MCP MCM, B-3001 Heverlee, Belgium
关键词
D O I
10.1049/el:19990731
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A general method to compensate for differences in probe-tip-to-line geometry and substrate permittivity between measurements and the calibration wafer is presented. The propagation constant, characteristic impedance and parameters of a general lumped-element equivalent discontinuity model are extracted at each frequency point based on measurements of two lines with different lengths.
引用
收藏
页码:1087 / 1088
页数:2