Development of a monochromatic, uniform source facility for calibration of radiance and irradiance detectors from 0.2 μm to 12 μm

被引:8
作者
Lykke, KR [1 ]
Shaw, PS [1 ]
Hanssen, LM [1 ]
Eppeldauer, GP [1 ]
机构
[1] Natl Inst Stand & Technol, Div Opt Technol, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/36/2/8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A radiometric source facility is being constructed with narrowband, widely tuneable lasers covering the wavelength range approximate to 200 nm to 12 mu m. The laser beam will pass into an integrating sphere and the output will be large-area, highly uniform, and monochromatic. This facility will be used to make a wide variety of radiometric measurements, such as radiant power, radiance, irradiance, transmittance and reflectance.
引用
收藏
页码:141 / 146
页数:6
相关论文
共 18 条
[1]   HIGHLY STABLE, MONOCHROMATIC AND TUNABLE OPTICAL RADIATION SOURCE AND ITS APPLICATION TO HIGH-ACCURACY SPECTROPHOTOMETRY [J].
ANDERSON, VE ;
FOX, NP ;
NETTLETON, DH .
APPLIED OPTICS, 1992, 31 (04) :536-545
[2]   Sum-frequency generation of continuous-wave light at 194 nm [J].
Berkeland, DJ ;
Cruz, FC ;
Bergquist, JC .
APPLIED OPTICS, 1997, 36 (18) :4159-4162
[3]   93% pump depletion, 3.5-W continuous-wave, singly resonant optical parametric oscillator [J].
Bosenberg, WR ;
Drobshoff, A ;
Alexander, JI ;
Myers, LE ;
Byer, RL .
OPTICS LETTERS, 1996, 21 (17) :1336-1338
[4]   Ultra-violet light generation at 205 nm by two frequency doubling steps of a cw titanium-sapphire laser [J].
Bourzeix, S ;
deBeauvoir, B ;
Nez, F ;
deTomasi, F ;
Julien, L ;
Biraben, F .
OPTICS COMMUNICATIONS, 1997, 133 (1-6) :239-244
[5]   EFFICIENT FREQUENCY DOUBLING OF A CONTINUOUS-WAVE TITANIUM - SAPPHIRE LASER IN AN EXTERNAL ENHANCEMENT CAVITY [J].
BOURZEIX, S ;
PLIMMER, MD ;
NEZ, F ;
JULIEN, L ;
BIRABEN, F .
OPTICS COMMUNICATIONS, 1993, 99 (1-2) :89-94
[6]   Near infrared radiometer standards [J].
Eppeldauer, G .
OPTICAL RADIATION MEASUREMENTS III, 1996, 2815 :42-54
[7]  
FOWLER JB, 1979, 987 NBS
[8]   High-accuracy characterisation and applications of filter radiometers [J].
Fox, NP ;
Chunnilall, CJ ;
Harrison, NJ ;
Hartree, WS .
OPTICAL RADIATION MEASUREMENTS III, 1996, 2815 :32-41
[9]   National Institute of Standards and Technology high-accuracy cryogenic radiometer [J].
Gentile, TR ;
Houston, JM ;
Hardis, JE ;
Cromer, CL ;
Parr, AC .
APPLIED OPTICS, 1996, 35 (07) :1056-1068
[10]   Realization of a scale of absolute spectral response using the national institute of standards and technology high-accuracy cryogenic radiometer [J].
Gentile, TR ;
Houston, JM ;
Cromer, CL .
APPLIED OPTICS, 1996, 35 (22) :4392-4403