Performance evaluation of the Sysmex haematology XN modular system

被引:148
作者
Briggs, Carol [1 ]
Longair, Ian [1 ]
Kumar, Punamar [1 ]
Singh, Deepak [2 ]
Machin, Samuel J. [1 ]
机构
[1] Univ Coll London Hosp, Dept Haematol, London W1T 4EU, England
[2] Doctors Lab, Dept Haematol, London, England
关键词
IMMATURE PLATELET FRACTION; XE-2100; SCHEME;
D O I
10.1136/jclinpath-2012-200930
中图分类号
R36 [病理学];
学科分类号
100103 [病原生物学];
摘要
Background The Sysmex XN haematology instrument performs automatic reflex testing, depending on sample results. A nucleated red blood cell (NRBC) count is provided on all samples. The instrument has a smaller footprint (34%) than previous Sysmex XE analysers. Methods An evaluation comparing all results to the Sysmex XE-2100 and manual microscopic differential and morphology (n = 390) was performed followed by a workflow study of 1000 samples to compare speed of operation and number of blood films reviews required from both systems. Results The new features on the instrument are: (1) white cell and NRBC channel, all samples include the NRBC count; (2) white cell precursor channel: false positive flags for blasts, abnormal lymphocytes and atypical lymphocytes are reduced significantly without a statistical increase of false negatives; (3) low white cell count mode: suggested setting of <0.5 x 10(9)/l. An extended count is more precise and provides an accurate differential. Fluorescent platelet count is performed in a dedicated channel. If the red cell or platelet size histograms are abnormal or if the platelet count is low, then a fluorescent platelet count is automatically performed. Good correlation with the XE-2100 and manual differential was found and the improved results compared to the reference flow cytometric analysis for platelet counts, especially below 30 x 10(9)/l (XE-2100, R-2 = 0.500; XN, R-2 = 0.875). Conclusion The XN showed reduced sample turnaround time of 10% and reduced number of blood films for examination, 49% less than the XE-2100 without loss of sensitivity with more precise and accurate results on low cell counts.
引用
收藏
页码:1024 / 1030
页数:7
相关论文
共 19 条
[1]
Barnes P W, 2005, Lab Hematol, V11, P83, DOI 10.1532/LH96.05019
[2]
Assessment of an immature platelet fraction (IPF) in peripheral thrombocytopenia [J].
Briggs, C ;
Kunka, S ;
Hart, D ;
Oguni, S ;
Machin, SJ .
BRITISH JOURNAL OF HAEMATOLOGY, 2004, 126 (01) :93-99
[3]
New quantitative parameters on a recently introduced automated blood cell counter -: the XE 2100™ [J].
Briggs, C ;
Harrison, P ;
Grant, D ;
Staves, J ;
Machin, SJ .
CLINICAL AND LABORATORY HAEMATOLOGY, 2000, 22 (06) :345-350
[4]
REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[5]
Clinical and Laboratory Standardization Institute, 2007, H20A2 CLSI
[6]
The Accuracy of Platelet Counting in Thrombocytopenic Blood Samples Distributed by the UK National External Quality Assessment Scheme for General Haematology [J].
De la Salle, Barbara J. ;
McTaggart, Paul N. ;
Briggs, Carol ;
Harrison, Paul ;
Dore, Caroline J. ;
Longair, Ian ;
Machin, Samuel J. ;
Hyde, Keith .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 137 (01) :65-74
[7]
ENGLAND JM, 1994, CLIN LAB HAEMATOL, V16, P157
[8]
Automated enumeration of immature Granulocytes [J].
Fernandes, Bernard ;
Hamaguchi, Yukio .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2007, 128 (03) :454-463
[9]
Stability of hematological analytes depends on the hematology analyser used: A stability study with Bayer Advia 120, Beckman Coulter LH 750 and Sysmex XE 2100 [J].
Imeri, Fatime ;
Herklotz, Roberto ;
Risch, Lorenz ;
Arbetsleitner, Christine ;
Zerlauth, Manfred ;
Risch, Gerhard M. ;
Huber, Andreas R. .
CLINICA CHIMICA ACTA, 2008, 397 (1-2) :68-71
[10]
The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755 [J].
Jean, A. ;
Boutet, C. ;
Lenormand, B. ;
Callat, M. -P. ;
Buchonnet, G. ;
Barbay, V. ;
Basuyau, J. -P. ;
Vasse, M. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2011, 33 (02) :138-145