The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755

被引:33
作者
Jean, A. [2 ]
Boutet, C. [2 ]
Lenormand, B. [2 ]
Callat, M. -P. [2 ]
Buchonnet, G. [2 ]
Barbay, V. [2 ]
Basuyau, J. -P. [3 ]
Vasse, M. [1 ,2 ]
机构
[1] CHU Charles Nicolle, Lab Hematol Biol, Inst Biol Clin, F-76031 Rouen, France
[2] CHU Charles Nicolle, IHU, Rouen, France
[3] Ctr Henri Becquerel, Biol Lab, F-76038 Rouen, France
关键词
Performance evaluation; complete blood count; differential; flow cytometry; haematology analyzer DxH 800; SYSMEX XE-2100; PLATELET;
D O I
10.1111/j.1751-553X.2010.01257.x
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
P>Introduction: The analytical performance and the abnormality messages on differential (flags) of the new analyzer Beckman Coulter DxH 800 were compared with those of the LH 755. Methods: First, we evaluated the accuracy of the results of the DxH 800, in comparison with the LH 755, in 125 samples without alarm using unflagged sample results on both analyzers. Second, flagged samples on the LH 755 but not flagged by the DxH 800 were evaluated by flow cytometry for accuracy of the DxH 800 results. Finally, we evaluated the sensitivity and specificity of abnormality messages on differential given by the analyzers, in comparison with manual blood smears. Results: The correlation coefficients (R) for complete blood count parameters and differential demonstrated that the DxH 800 results were similar to that of LH 755. Excellent correlation coefficients between DxH 800 and flow cytometry results were found for white blood cell count (R = 0.985, n = 31), platelet count (R = 0.976, n = 51) and nucleated red blood cells (R = 0.966, n = 37). The overall performance showed an increased sensitivity (0.892) and specificity (0.864) of the flags on DxH 800 when compared to the LH 755 (0.846 and 0.733, respectively). Conclusion: The DxH 800 provides reliable results and increases laboratory efficiency by reducing working time and costs associated with the optical validation of the results.
引用
收藏
页码:138 / 145
页数:8
相关论文
共 10 条
[1]
Performance evaluation of the Sysmex XS-1000i automated haematology analyser [J].
Ghys, T. ;
Malfait, R. ;
Van den Bossche, J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (05) :560-566
[2]
Harrison P, 2001, AM J CLIN PATHOL, V115, P448
[3]
Diagnostic performance of the variant lymphocyte flag of the Abbott Cell-Dyn 4000 haematology analyser [J].
Hoffmann, JJML ;
Hoedemakers, RMJ .
CLINICAL AND LABORATORY HAEMATOLOGY, 2004, 26 (01) :9-13
[4]
Hubl W, 1996, CYTOMETRY, V26, P69, DOI 10.1002/(SICI)1097-0320(19960315)26:1<69::AID-CYTO11>3.0.CO
[5]
2-Q
[6]
Evaluation of the coulter LH 750 haematology analyzer compared with flow cytometry as the reference method for WBC, platelet and nucleated RBC count [J].
Igout, J ;
Fretigny, M ;
Vasse, M ;
Callat, MP ;
Silva, M ;
Willemont, L ;
Gelle, M ;
Lenormand, B .
CLINICAL AND LABORATORY HAEMATOLOGY, 2004, 26 (01) :1-7
[7]
Comparison of four hematology analyzers, CELL-DYN Sapphire, ADVIA 120, Coulter LH 750, and Sysmex XE-2100, in terms of clinical usefulness [J].
Kang, S. -H. ;
Kim, H. K. ;
Ham, C. K. ;
Lee, D. S. ;
Cho, H. -I. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2008, 30 (06) :480-486
[8]
Performance evaluation of Abbott CELL-DYN Ruby for routine use [J].
Lehto, T. ;
Hedberg, P. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2008, 30 (05) :400-407
[9]
Preliminary evaluation of the new hematology analyzer COULTER® GEN•STM in a university hospital [J].
Picard, F ;
Gicquel, C ;
Marnet, L ;
Guesnu, M ;
Levy, JP .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1999, 37 (06) :681-686
[10]
Ruzicka K, 2001, ARCH PATHOL LAB MED, V125, P391