Performance evaluation of the Sysmex XS-1000i automated haematology analyser

被引:20
作者
Ghys, T. [1 ]
Malfait, R. [1 ]
Van den Bossche, J. [1 ]
机构
[1] AZ Middelheim, Clin Lab Middelheim, B-2020 Antwerp, Belgium
关键词
Performance evaluation; haematology analyser; complete blood count; differential; XS-1000i; SODIUM LAURYL SULFATE; HEMOGLOBIN DETERMINATION; DIFFERENTIAL LEUKOCYTE; COUNT;
D O I
10.1111/j.1751-553X.2008.01081.x
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
P>The Sysmex((R)) XS-1000i is a compact new, fully automated haematology analyser, designed to generate complete blood counts with five-part leucocyte differential. In our study, a Sysmex XS-1000i instrument was evaluated according to Clinical Laboratory Standards Institute (CLSI) and International Council for Standardization in Haematology (ICSH) guidelines. Precision, carry-over and linearity were determined. Using a total of 700 patient samples, results from the Sysmex XS-1000i were compared with those from a Sysmex XE-2100, an Abbott Cell Dyn 4000 and the manual reference leucocyte differential. Using quality control material, total and within-run imprecision was less than 3% except for platelets. The system demonstrated good linearity over the entire reporting range and no carry-over (< 0.5%). The Sysmex XS-1000i showed good correlation with XE-2100, CD-4000 and the manual reference leucocyte differential. Overall flagging sensitivity and specificity were 91% and 48%, respectively. In conclusion, the Sysmex XS-1000i demonstrated good analytical performance, is able to generate a complete blood count with five-part differential on low blood volumes and has considerable back-up capacity.
引用
收藏
页码:560 / 566
页数:7
相关论文
共 17 条
[1]
Aulesa C, 2006, Lab Hematol, V12, P201, DOI 10.1532/LH96.06016
[2]
Continuing developments with the automated platelet count [J].
Briggs, C. ;
Harrison, P. ;
Machin, S. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2007, 29 (02) :77-91
[3]
*CLSI, 2007, APPR STAND 2 ED H20A
[4]
*CLSI, 2004, APPR GUID 2 ED EP5 A
[5]
ENGLAND JM, 1994, CLIN LAB HAEMATOL, V16, P157
[6]
Between-examiner reproducibility in manual differential leukocyte counting [J].
Fuentes-Arderiu, Xavier ;
Garcia-Panyella, Mariona ;
Dot-Bach, Dolors .
ACCREDITATION AND QUALITY ASSURANCE, 2007, 12 (12) :643-645
[7]
Evaluation of the Abbott CELL-DYN 4000 hematology analyzer [J].
Grimaldi, E ;
Scopacasa, F .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 113 (04) :497-505
[8]
KANG SH, 2007, INT J LAB HEMATOLOGY
[9]
AN EVALUATION OF HEMOGLOBIN DETERMINATION USING SODIUM LAURYL SULFATE [J].
KARSAN, A ;
MACLAREN, I ;
CONN, D ;
WADSWORTH, L .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1993, 100 (02) :123-126
[10]
Possibilities and limitations of automated platelet counting procedures in the thrombocytopenic range [J].
Kunz, D .
SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2001, 27 (03) :229-235