Continuing developments with the automated platelet count

被引:149
作者
Briggs, C.
Harrison, P.
Machin, S. J.
机构
[1] UCL Hosp, Dept Haematol, London W1T 4EU, England
[2] Churchill Hosp, Oxford Haemophilia Ctr, Oxford, England
[3] Churchill Hosp, Thrombosis Unit, Oxford, England
关键词
platelets; automated counting; problems in platelet counting; thrombocytopenia; platelet transfusion;
D O I
10.1111/j.1751-553X.2007.00909.x
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
The four main procedures for platelet counting are: manual phase contrast microscopy, impedance, optical light scatter/fluorescence and flow cytometry. Early methods to enumerate platelets were inaccurate and irreproducible. The manual count is still recognized as the gold standard or reference method, and until very recently the calibration of platelet counts by the manufacturers of automated cell counters and it is quality control material was performed by this method. However, time-consuming and results in high levels of imprecision. The introduction of automated full blood counters using impedance technology resulted in a dramatic improvement in precision. However, impedance counts still have limitations as cell size analysis cannot discriminate platelets from other similar-sized particles. More recently, light scatter or fluorescence methods have been introduced for automated platelet counting, but there are still occasional cases where an accurate platelet count remains a challenge. Thus, there has been interest in the development of an improved reference procedure to enable optimization of automated platelet counting. This method utilizes monoclonal antibodies to platelet cell surface antigens conjugated to a suitable fluorophore. This permits the possible implementation of a new reference method to calibrate cell counters, assign values to calibrators, and to obtain a direct platelet count on a variety of pathological samples. in future, analysers may introduce additional platelet parameters; a reliable method to quantify immature or reticulated platelets would be useful.
引用
收藏
页码:77 / 91
页数:15
相关论文
共 44 条
[1]
A simple technique to determine thrombopoiesis level using immature platelet fraction (IPF) [J].
Abe, Yasunori ;
Wada, Hideo ;
Tomatsu, Hiroaki ;
Sakaguchi, Akane ;
Nishioka, Junji ;
Yabu, Yasunori ;
Onishi, Katsuya ;
Nakatani, Kaname ;
Morishita, Yoshitaka ;
Oguni, Shinichiro ;
Nobori, Tsutomu .
THROMBOSIS RESEARCH, 2006, 118 (04) :463-469
[2]
Trigger factors for prophylactic platelet transfusion [J].
Ancliff, PJ ;
Machin, SJ .
BLOOD REVIEWS, 1998, 12 (04) :234-238
[3]
[Anonymous], 1996, LAB HAEMATOL
[4]
AULT KA, 1997, LAB HEMATOL, V3, P125
[5]
CLINICAL-EVALUATION OF A NEW ELECTRICAL-IMPEDANCE INSTRUMENT FOR COUNTING PLATELETS IN WHOLE-BLOOD [J].
BACUS, JW ;
WATT, S ;
TROBAUGH, FE .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1980, 73 (05) :655-663
[6]
A PARALLEL EVALUATION OF 4 AUTOMATED HEMATOLOGY ANALYZERS [J].
BENTLEY, SA ;
JOHNSON, A ;
BISHOP, CA .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1993, 100 (06) :626-632
[7]
Bizzozero G, 1882, VIRCHOWS ARCH PATHOL, V90, P261
[8]
Platelet size has diagnostic predictive value in patients with thrombocytopenia [J].
Bowles, KM ;
Cooke, LJ ;
Richards, EM ;
Baglin, TP .
CLINICAL AND LABORATORY HAEMATOLOGY, 2005, 27 (06) :370-373
[9]
BRECHER G, 1953, AM J CLIN PATHOL, V23, P15
[10]
Immature platelet fraction measurement: a future guide to platelet transfusion requirement after haematopoietic stem cell transplantation [J].
Briggs, C ;
Hart, D ;
Kunka, S ;
Oguni, S ;
Machin, SJ .
TRANSFUSION MEDICINE, 2006, 16 (02) :101-109