Application of a portable immunosensor to detect the explosives TNT and RDX in groundwater samples

被引:49
作者
Bart, JC [1 ]
Judd, LL [1 ]
Hoffman, KE [1 ]
Wilkins, AM [1 ]
Kusterbeck, AW [1 ]
机构
[1] USN,RES LAB,CTR BIOMOL SCI & ENGN,BIOSENSORS & BIOMAT LAB,WASHINGTON,DC 20375
关键词
D O I
10.1021/es960777l
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In the summer of 1995, the EPA conducted a performance evaluation of several commercially available test kits and two prototype biosensor devices to determine if any of these technologies were feasible as an alternative to the standard EPA SW-846 Method 8330 test for explosives in aqueous samples. These on-site assays offer potential advantages in cost, assay time, and convenience over the traditional analysis via high-performance liquid chromatography (HPLC). The continuous flow immunosensor (CFI) was one of the biosensors participating in this trial, which took place at two Superfund sites located on military bases. The CFI uses a small column filled with plastic beads containing immobilized antibodies against the explosive being assayed and a fluorescent dye-labeled explosive analog. Detection occurs when the native explosive in the sample is swept into the column and displaces some of the dye-labeled analog, which is quantified via a fluorometer. Results from these tests showed that the CFI could produce data comparable to HPLC with no significant problems with cross-reactivity of the antibodies against other explosives or their breakdown products.
引用
收藏
页码:1505 / 1511
页数:7
相关论文
共 40 条
[31]   DETERMINATION OF EXPLOSIVES AND THEIR BIODEGRADATION PRODUCTS IN CONTAMINATED SOIL AND WATER FROM FORMER AMMUNITION PLANTS BY AUTOMATED MULTIPLE DEVELOPMENT HIGH-PERFORMANCE THIN-LAYER CHROMATOGRAPHY [J].
STEUCKART, C ;
BERGERPREISS, E ;
LEVSEN, K .
ANALYTICAL CHEMISTRY, 1994, 66 (15) :2570-2577
[32]  
TWIBELL JD, 1984, J FORENSIC SCI, V29, P277
[33]  
*US EPA, 1988, HLTH ADV RDX
[34]  
*US EPA OFF DRINK, 1989, TRIN HLTH ADV
[35]  
Walsh M. E., 1993, 935 CRREL
[36]   CONTINUOUS-FLOW IMMUNOSENSOR FOR DETECTION OF EXPLOSIVES [J].
WHELAN, JP ;
KUSTERBECK, AW ;
WEMHOFF, GA ;
BREDEHORST, R ;
LIGLER, FS .
ANALYTICAL CHEMISTRY, 1993, 65 (24) :3561-3565
[37]   MASS-SPECTRAL FRAGMENTATION PATHWAYS IN SOME DINITROAROMATIC COMPOUNDS STUDIED BY COLLISION-INDUCED DISSOCIATION AND TANDEM MASS-SPECTROMETRY [J].
YINON, J .
ORGANIC MASS SPECTROMETRY, 1992, 27 (06) :689-694
[38]  
YINON J, 1993, MODERN METHODS APPLA
[39]  
Yinon J, 1990, Toxicity and metabolism of explosives
[40]   DETERMINATION OF 2,4,6-TRINITROTOLUENE AND ITS BIODEGRADATION PRODUCTS BY NORMAL-PHASE LIQUID-CHROMATOGRAPHY [J].
ZOU, HF ;
ZHOU, SF ;
HU, X ;
HONG, MF ;
ZHANG, YK ;
LU, PC .
ANALYTICA CHIMICA ACTA, 1994, 291 (1-2) :205-210