Resistance of the lumbar spine against axial compression forces after implantation of three different posterior lumbar interbody cages

被引:31
作者
Krammer, M
Dietl, R
Lumenta, CB
Kettler, A
Wilke, HJ
Büttner, A
Claes, L
机构
[1] Tech Univ Munich, Acad Hosp Munchen Bogenhausen, Dept Neurosurg, D-81925 Munich, Germany
[2] Univ Ulm, Inst Orthoped Res & Biomech, Ulm, Germany
[3] Univ Munich, Inst Legal Med, Munich, Germany
关键词
lumbar spine; interbody fusion; compressive behaviour; cyclic testing;
D O I
10.1007/s007010100017
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background. The aim of using interbody fusion cages is to distract the degeneratively decreased disc height to decompress the neural structures in the intervertebral foramina and allow bony fusion. Prerequisite for a successful fusion therapy is a high resistance against subsidence and breakage. Method. Three types of implants, a cylindrical threaded titanium cage (Ray(TM)) (1c), a bullet shaped PEEK cage (Stryker(TM)) (1a) and a rectangular titanium cage with an endplate anchorage device (Marquardt(TM)) (1b) were implanted in eight monosegmental lumbar spine specimens (L 2/3 and L 4/5). Each specimen underwent a cyclic loading test with 40000 cycles at a rate of 5 Hz. A cyclic axial compression force ranging from 200 Newton [N] to 1000 N was applied and the axial translation recorded simultaneously to determine the subsidence tendency. After this procedure the specimens were tested with a progressive axial force until breakage. Findings. There were only small differences in the subsidence tendency for the three cage designs. The height reduction due to cyclic loading ranged between 0.9 mm (Marquardt(TM)), 1.2 mm (Stryker(TM)) and 1.4 mm (Ray(TM)). The median break force ranged from 5486 N (Marquardt(TM)), 8359 N (Stryker(TM)) to 8413 N (Ray(TM)). No correlation between bone mineral density and failure load could be detected. Interpretation. Endplate preparation and cage design of the tested implants do not seem to influence the resistance of the segment against cyclic axial compression. The compression with a continuously increasing load revealed that an implant-bone failure is not to be expected in physiological limits for all three cage types.
引用
收藏
页码:1217 / 1222
页数:6
相关论文
共 24 条
[1]
BAGBY GW, 1987, P ANN M N AM SPIN SO, P89
[2]
A CARBON-FIBER IMPLANT TO AID INTERBODY LUMBAR FUSION - 2-YEAR CLINICAL-RESULTS IN THE 1ST 26 PATIENTS [J].
BRANTIGAN, JW ;
STEFFEE, AD .
SPINE, 1993, 18 (14) :2106-2117
[3]
A CARBON-FIBER IMPLANT TO AID INTERBODY LUMBAR FUSION - MECHANICAL TESTING [J].
BRANTIGAN, JW ;
STEFFEE, AD ;
GEIGER, JM .
SPINE, 1991, 16 (06) :S277-S282
[4]
DEFORMATION OF THE VERTEBRAL ENDPLATE UNDER AXIAL LOADING OF THE SPINE [J].
BRINCKMANN, P ;
FROBIN, W ;
HIERHOLZER, E ;
HORST, M .
SPINE, 1983, 8 (08) :851-856
[6]
DEBOWES RM, 1984, AM J VET RES, V45, P191
[7]
COMPARISON OF DISK SPACE HEIGHTS AFTER ANTERIOR LUMBAR INTERBODY FUSION [J].
DENNIS, S ;
WATKINS, R ;
LANDAKER, S ;
DILLIN, W ;
SPRINGER, D .
SPINE, 1989, 14 (08) :876-878
[8]
FAILURE MECHANISMS IN HUMAN VERTEBRAL CANCELLOUS BONE [J].
FYHRIE, DP ;
SCHAFFLER, MB .
BONE, 1994, 15 (01) :105-109
[9]
Holte DC, 1984, EUR SPINE J, V3, P32
[10]
JASLOW IA, 1946, SURG GYNECOL OBSTET, V82, P215