Acetabular Loosening Using an Extended Offset Polyethylene Liner

被引:22
作者
Archibeck, Michael J. [1 ]
Cummins, Tamara [1 ]
Junick, Daniel W. [1 ]
White, Richard E., Jr. [1 ]
机构
[1] New Mexico Orthopaed, New Mexico Ctr Joint Replacement Surg, Albuquerque, NM 87106 USA
关键词
COMPONENTS; FIXATION;
D O I
10.1007/s11999-008-0479-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
The use of extended offset femoral components and acetabular liners helps restore preoperative offset during hip arthroplasty. We report a relatively high acetabular component aseptic loosening rate with the use of offset polyethylene liners. We reviewed 1919 primary and 346 revision total hip arthroplasties (THAs). A 7-mm offset acetabular liner was used in 120 of the primary and 100 of the revision THAs. The aseptic loosening rate in the primary THA group was 0.12% in the standard offset and 4.2% in the extended offset groups at a minimum of 2 years (mean, 3.6 years; range, 2-9 years) followup. The aseptic loosening rate in the revision group was 1.7% in the standard and 7% in the extended offset groups at a mean of 4 years (range, 2-9 years) followup. Although extended offset acetabular liners help restore hip offset, torsional force applied to the implant-bone interface may have a detrimental effect on fixation. We found a relatively high failure rate in our primary and revision acetabular components used with an offset liner. Level of Evidence: Level III, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.
引用
收藏
页码:188 / 193
页数:6
相关论文
共 6 条
[1]
THE INITIAL STABILITY OF UNCEMENTED ACETABULAR COMPONENTS [J].
CURTIS, MJ ;
JINNAH, RH ;
WILSON, VD ;
HUNGERFORD, DS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :372-376
[2]
Offset acetabular components introduce torsion on the implant and may increase the risk of fixation failure [J].
Giori, NJ .
JOURNAL OF ARTHROPLASTY, 2003, 18 (01) :89-91
[3]
HURWITZ DE, 1998, ADULT HIP, V1, P80
[4]
The initial fixation of the press-fit acetabular shell - clinical observation and experimental study [J].
Kaneko, K ;
Inoue, Y ;
Yanagihara, Y ;
Uta, S ;
Mogami, A ;
Iwase, H .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2000, 120 (5-6) :323-325
[5]
Lawless JF., 2002, STAT MODELS METHODS
[6]
Acetabular revision with a 7-mm lateral offset component inserted without cement in patients who have acetabular deficiency [J].
McGrory, BJ .
JOURNAL OF ARTHROPLASTY, 2005, 20 (04) :443-453