Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, Ocular Response Analyzer, and Goldmann tonometry in glaucomatous eyes

被引:58
作者
Annette, Hager [1 ]
Kristina, Loge [1 ]
Bernd, Schroeder [1 ]
Mark-Oliver, Fuellhas [1 ]
Wolfgang, Wiegand [1 ]
机构
[1] Asklepios Klin Nord Heidberg, Dept Ophthalmol, D-22419 Hamburg, Germany
关键词
Goldmann applanation tonometry; dynamic contour tonometry; Ocular Response Analyzer; viscoelastic properties of the cornea; corneal hysteresis;
D O I
10.1097/IJG.0b013e31815c3ad3
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
Purpose: The dynamic contour tonometer (DCT; Pascal tonometer) and the Ocular Response Analyzer (ORA) are novel tonometers designed to measure intraocular pressure (IOP) independent of corneal properties and central corneal thickness (CCT), respectively. We wanted to compare the corneal compensated IOP (IOPcc) as measured by ORA with IOP values measured by DCT and Goldmann applanation tonometry (GAT) with respect to IOP readings and the influence of corneal hysteresis (CH) and CCT in glaucoma patients. Patients and Methods: In a Study of 94 glaucomatous eyes, IOP Measurements by ORA, DCT, and GAT were compared, and the effects of CCT and CH were analyzed. All measurements were taken by 1 of the authors only. Results: The Mean CCT was 550 +/- 44 mu m and mean CH 10.24 +/- 3.3 mm Hg. The mean value for IOPcc was 17.94 +/- 5.9 mm Hg, 15.14 +/- 3.7 mm Hg for DCT, and 14.3 +/- 4.3 mm Hg for GAT. The mean difference was 3.65 +/- 3.85 mm Hg between IOPcc and GAT and 2.80 +/- 4.9 mm Hg between IOPcc and DCT (P < 0.001) and -0.85 +/- 3.3 mm Hg between GAT and DCT (P = 0.015). None of the methods was related to CCT. In contrast, CH was highly statistically significant related to IOPcc (P < 0.0001). whereas GAT and DCT showed only weak relation to CH (P = 0.05). Conclusions: IOP measurements with the ORA are significantly higher than DCT or GAT with no apparent linear correction factor. To date, we cannot estimate the impact of differences in CH in relation to IOP and its measurement.
引用
收藏
页码:361 / 365
页数:5
相关论文
共 32 条
[1]
Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[2]
Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS) [J].
Brandt, JD ;
Beiser, JA ;
Kass, MA ;
Gordon, MO .
OPHTHALMOLOGY, 2001, 108 (10) :1779-1788
[3]
Reduction in intraocular pressure after excimer laser photorefractive keratectomy - Correlation with pretreatment myopia [J].
Chatterjee, A ;
Shah, S ;
Bessant, DA ;
Naroo, SA ;
Doyle, SJ .
OPHTHALMOLOGY, 1997, 104 (03) :355-359
[4]
RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES [J].
DIELEMANS, I ;
VINGERLING, JR ;
HOFMAN, A ;
GROBBEE, DE ;
DEJONG, PTVM .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1994, 232 (03) :141-144
[5]
Intraocular pressure and corneal thickness - A comparison between non-contact tonometry and applanation tonometry [J].
Domke, N. ;
Hager, A. ;
Wiegand, W. .
OPHTHALMOLOGE, 2006, 103 (07) :583-587
[6]
DRAEGER J, 1975, KLIN MONATSBL AUGENH, V167, P27
[7]
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[8]
Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings [J].
Feltgen, N ;
Leifert, D ;
Funk, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (01) :85-87
[9]
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[10]
Evaluation of tonometric correction factors [J].
Gunvant, P ;
Oleary, DJ ;
Baskaran, M ;
Broadway, DC ;
Watkins, RJ ;
Vijaya, L .
JOURNAL OF GLAUCOMA, 2005, 14 (05) :337-343