A freehand elastographic imaging approach for clinical breast imaging: System development and performance evaluation

被引:127
作者
Doyley, MM [1 ]
Bamber, JC [1 ]
Fuechsel, F [1 ]
Bush, NL [1 ]
机构
[1] Canc Res Inst, Sutton, Surrey, England
基金
英国医学研究理事会;
关键词
breast imaging; freehand elastography; hand-induced transducer motion; tissue elasticity; tumour detection;
D O I
10.1016/S0301-5629(01)00429-X
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A prototype freehand elastographic imaging system has been developed for clinical breast imaging. The system consists of a fast data acquisition system, which is able to capture sequences of intermediate frequency echo frames at full frame rate from a commercial ultrasound scanner whilst the breast is deformed using hand-induced transducer motion. Two-dimensional echo tracking was used in combination with global distortion compensation and multi-compression averaging to minimise decorrelation noise incurred when stress is applied using hand-induced transducer motion. Experiments were conducted on gelatine phantoms to evaluate the quality of elastograms produced using the prototype system relative to those produced using mechanically induced transducer motion. The strain sensitivity and contrast-to-noise ratio of freehand elastograms compared favourably with elastograms produced using mechanically induced transducer motion. However, better dynamic range and signal-to-noise ratio was achieved when elastograms were created using mechanically induced transducer motion. Despite the loss in performance incurred when stress is applied using hand-induced transducer motion, it was concluded that the prototype system performed sufficiently well to warrant clinical evaluation. (C) 2001 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1347 / 1357
页数:11
相关论文
共 28 条
[1]   Elastographic axial resolution criteria: An experimental study [J].
Alam, SK ;
Ophir, J ;
Varghese, T .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (01) :304-309
[2]   QUANTITATIVE-EVALUATION OF REAL-TIME ULTRASOUND FEATURES OF THE BREAST [J].
BAMBER, JC ;
DEGONZALEZ, L ;
COSGROVE, DO ;
SIMMONS, P .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1988, 14 :81-87
[3]  
BAMBER JC, 1995, ACOUSTICAL IMAGING, V22
[4]  
Cespedes E I, 1997, Semin Interv Cardiol, V2, P55
[5]   REDUCTION OF IMAGE NOISE IN ELASTOGRAPHY [J].
CESPEDES, I ;
OPHIR, J .
ULTRASONIC IMAGING, 1993, 15 (02) :89-102
[6]   Testing the limitations of 2-D companding for strain imaging using phantoms [J].
Chaturvedi, P ;
Insana, MF ;
Hall, TJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (04) :1022-1031
[7]   2-D companding for noise reduction in strain imaging [J].
Chaturvedi, P ;
Insana, MF ;
Hall, TJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (01) :179-191
[8]   ULTRASOUND TISSUE DISPLACEMENT IMAGING WITH APPLICATION TO BREAST-CANCER [J].
CHEN, EJ ;
ADLER, RS ;
CARSON, PL ;
JENKINS, WK ;
OBRIEN, WD .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (09) :1153-1162
[9]   THE IMPACT OF VARIOUS IMAGING PARAMETERS ON ULTRASONIC DISPLACEMENT AND VELOCITY ESTIMATES [J].
CHEN, EJ ;
JENKINS, WK ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1994, 41 (03) :293-301
[10]   Characterization of plaque components with intravascular ultrasound elastography in human femoral and coronary arteries in vitro [J].
de Korte, CL ;
Pasterkamp, G ;
van der Steen, AFW ;
Woutman, HA ;
Bom, N .
CIRCULATION, 2000, 102 (06) :617-623