Use of a polybrene capillary coating in capillary electrophoresis for rapid analysis of hemoglobin variants with on-line detection via an ion trap storage reflectron time-of-flight mass spectrometer

被引:77
作者
Li, MX [1 ]
Liu, L [1 ]
Wu, JT [1 ]
Lubman, DM [1 ]
机构
[1] UNIV MICHIGAN,DEPT CHEM,ANN ARBOR,MI 48109
关键词
D O I
10.1021/ac970076m
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A polybrene capillary coating in capillary electrophoresis (CE) has been used for rapid analysis of hemoglobin variant digests, The use of the polybrene capillary coating has allowed sufficient separation to resolve the large number of digest products formed upon tryptic digestion of the whole protein, so that prior separation of the hemoglobin alpha and beta chains is not required, The resolution of the digest peaks obtained by CE is sufficient so that even single amino acid substitutions can easily be detected using UV absorption detection. The digest is further analyzed by capillary electrophoresis separation with on-line detection using electrospray ionization interfaced to the ion trap storage/reflectron time of flight device (CE/ESI-IT/reTOF), where a comparison of the total ion electropherograms and mass spectra of the mutant and normal hemoglobins can detect the presence of a mutation site. The CE separation and mass analysis can be accomplished in typically 10-15 min. The unique capability of the CE/ESI-IT/reTOF system for detection of fast separations with narrow peaks that may be under 1 s fwhm is demonstrated. The speed of this system is essential for resolution of the large number of peaks that are separated in a short time duration using CE separations.
引用
收藏
页码:2451 / 2456
页数:6
相关论文
共 34 条
[1]  
BASSET P, 1978, BLOOD, V51, P971
[2]  
Bateman KP, 1997, RAPID COMMUN MASS SP, V11, P307
[3]   CAPILLARY ELECTROPHORESIS MASS-SPECTROMETRY [J].
CAI, JY ;
HENION, J .
JOURNAL OF CHROMATOGRAPHY A, 1995, 703 (1-2) :667-692
[4]   THE USE OF CAPILLARY ELECTROPHORESIS FOR THE DETERMINATION OF HEMOGLOBIN-VARIANTS [J].
CASTAGNOLA, M ;
MESSANA, I ;
CASSIANO, L ;
RABINO, R ;
ROSSETTI, DV ;
GIARDINA, B .
ELECTROPHORESIS, 1995, 16 (08) :1492-1498
[5]   RAPID ANALYSIS OF LABELED GLOBIN CHAINS WITHOUT ACETONE PRECIPITATION OR DIALYSIS BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY AND ION-EXCHANGE CHROMATOGRAPHY [J].
CONGOTE, LF ;
KENDALL, AG .
ANALYTICAL BIOCHEMISTRY, 1982, 123 (01) :124-132
[6]   MICROSCALE IMMOBILIZED PROTEASE REACTOR COLUMNS FOR PEPTIDE-MAPPING BY LIQUID-CHROMATOGRAPHY MASS-SPECTRAL ANALYSES [J].
DAVIS, MT ;
LEE, TD ;
RONK, M ;
HEFTA, SA .
ANALYTICAL BIOCHEMISTRY, 1995, 224 (01) :235-244
[7]   CAPILLARY ZONE ELECTROPHORESIS AND MASS-SPECTROMETRY FOR THE CHARACTERIZATION OF GENETIC-VARIANTS OF HUMAN HEMOGLOBIN [J].
FERRANTI, P ;
MALORNI, A ;
PUCCI, P ;
FANALI, S ;
NARDI, A ;
OSSICINI, L .
ANALYTICAL BIOCHEMISTRY, 1991, 194 (01) :1-8
[8]   Protein identification by capillary zone electrophoresis/microelectrospray ionization-tandem mass spectrometry at the subfemtomole level [J].
Figeys, D ;
vanOostveen, I ;
Ducret, A ;
Aebersold, R .
ANALYTICAL CHEMISTRY, 1996, 68 (11) :1822-1828
[9]   CAPILLARY ELECTROPHORESIS [J].
GORDON, MJ ;
HUANG, XH ;
PENTONEY, SL ;
ZARE, RN .
SCIENCE, 1988, 242 (4876) :224-228
[10]  
HUISMAN THJ, 1986, HEMOGLOBINOPATHIES