Validation of the Microlife BP3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol

被引:63
作者
Cuckson, AC [1 ]
Reinders, A [1 ]
Shabeeh, H [1 ]
Shennan, AH [1 ]
机构
[1] St Thomas Hosp, GKT Sch Med, Maternal & Fetal Res Unit, London SE1 7EH, England
关键词
self-measurement; blood pressure; British Hypertension Society; Association for the Advancement of Medical Instrumentation; oscillometric;
D O I
10.1097/00126097-200212000-00005
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The market for devices for the self-measurement of blood pressure is growing, and as accuracy is of prime importance, there is increasing pressure for manufacturers to provide evidence of independent testing. Recent reviews have shown that only five automated upper arm devices for self-measurement of blood pressure have been recommended for use. We tested the Microlife BP 3BTO-A, a lightweight, upper arm, automated oscillometric device, according to a modified version of the British Hypertension Society protocol and also analysed the computer-generated oscillograms for possible causes of inaccuracy. Methods One hundred and twenty-six subjects were recruited from general medical and specialist clinics and from amongst the staff at Guy's and St Thomas' Hospital, London, UK. Only 85 of these were included in the final analysis. Nine sequential readings were taken by two trained observers alternating between the mercury sphygmomanometer and the device. The last seven readings were analysed according to the British Hypertension Society protocol. Modifications to the protocol were: (1) the exclusion of patients whose blood pressure varied by more than 15 mmHg between sequential observer readings and (2) limited testing in the low systolic pressure range. Results The Microlife achieved a grade A for both systolic and diastolic pressure according to the British Hypertension Society protocol. The mean differences (standard deviation) between the observers and the device were -1.6 (7.7) mmHg and -2.1 (6.3) mmHg for systolic and diastolic blood pressure, respectively, therefore also fulfilling the criteria set by the Association for the Advancement of Medical Instrumentation. Sub-analysis for different pressure ranges showed that the device was less accurate in the high-pressure range (> 160/100 mmHg). Conclusion The Microlife can be recommended for clinical use in an adult population. (C) 2002 Lippincott Williams Wilkins.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 10 条
[1]  
Association for the Advancement of Medical Instrumentation, 1993, SP101992 ANSIAAMI
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
Bortolotto L A, 1999, Blood Press Monit, V4, P21, DOI 10.1097/00126097-199904010-00004
[4]   CHARACTERIZATION OF THE OSCILLOMETRIC METHOD FOR MEASURING INDIRECT BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, M ;
COMBS, C ;
REINER, D ;
BABBS, CF .
ANNALS OF BIOMEDICAL ENGINEERING, 1982, 10 (06) :271-280
[5]   A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor [J].
Jones, CR ;
Taylor, K ;
Chowienczyk, P ;
Poston, L ;
Shennan, AH .
BLOOD PRESSURE MONITORING, 2000, 5 (04) :233-238
[6]  
Ng Kim-Gau, 1994, Journal of Clinical Engineering, V19, P452
[7]   Validation up-date [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2001, 6 (06) :275-280
[8]   State of the market in 2001 for blood pressure measuring devices [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2001, 6 (04) :171-176
[9]   Blood pressure measuring devices: recommendations of the European Society of Hypertension [J].
O'Brien, E ;
Waeber, B ;
Parati, G ;
Staessen, J ;
Myers, MG .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7285) :531-536
[10]  
O'Brien E., 1993, J HYPERTENSION S2, V11, P43, DOI DOI 10.1097/00004872-199306000-00013