A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology

被引:337
作者
Erlandsson, Kjell [1 ]
Buvat, Irene [2 ]
Pretorius, P. Hendrik [3 ]
Thomas, Benjamin A. [1 ]
Hutton, Brian F. [1 ]
机构
[1] UCL, Inst Nucl Med, London, England
[2] CNRS, UMR 8165, IMNC, F-91406 Orsay, France
[3] Univ Massachusetts, Sch Med, Dept Radiol, Div Nucl Med, Worcester, MA USA
基金
英国工程与自然科学研究理事会;
关键词
STANDARDIZED UPTAKE VALUE; MYOCARDIAL BLOOD-FLOW; MESOLIMBIC DOPAMINE TRANSMISSION; OF-INTEREST EVALUATION; POSITRON-EMISSION; IMAGE-RECONSTRUCTION; TUMOR RESPONSE; MONTE-CARLO; FDG PET; EXPECTATION MAXIMIZATION;
D O I
10.1088/0031-9155/57/21/R119
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Accurate quantification in PET and SPECT requires correction for a number of physical factors, such as photon attenuation, Compton scattering and random coincidences (in PET). Another factor affecting quantification is the limited spatial resolution. While considerable effort has gone into development of routine correction techniques for the former factors, less attention has been paid to the latter. Spatial resolution-related effects, referred to as 'partial volume effects' (PVEs), depend not only on the characteristics of the imaging system but also on the object and activity distribution. Spatial and/or temporal variations in PVE can often be confounding factors. Partial volume correction (PVC) could in theory be achieved by some kind of inverse filtering technique, reversing the effect of the system PSF. However, these methods are limited, and usually lead to noise-amplification or image artefacts. Some form of regularization is therefore needed, and this can be achieved using information from co-registered anatomical images, such as CT or MRI. The purpose of this paper is to enhance understanding of PVEs and to review possible approaches for PVC. We also present a review of clinical applications of PVC within the fields of neurology, cardiology and oncology, including specific examples.
引用
收藏
页码:R119 / R159
页数:41
相关论文
共 167 条
[1]  
Alessio A, 2008, NUCL SCI S IEEE, P3637
[2]   Application and Evaluation of a Measured Spatially Variant System Model for PET Image Reconstruction [J].
Alessio, Adam M. ;
Stearns, Charles W. ;
Tong, Shan ;
Ross, Steven G. ;
Kohlmyer, Steve ;
Ganin, Alex ;
Kinahan, Paul E. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (03) :938-949
[3]   Minimum cross-entropy reconstruction of PET images using prior anatomical information [J].
Ardekani, BA ;
Braun, M ;
Hutton, BF ;
Kanno, I ;
Iida, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2497-2517
[4]  
Avril N, 1997, J NUCL MED, V38, P1186
[5]   Evaluation of anatomy based reconstruction for partial volume correction in brain FDG-PET [J].
Baete, K ;
Nuyts, J ;
Van Laere, K ;
Van Paesschen, W ;
Ceyssens, S ;
De Ceuninck, L ;
Gheysens, O ;
Kelles, A ;
Van den Eynden, J ;
Suetens, P ;
Dupont, P .
NEUROIMAGE, 2004, 23 (01) :305-317
[6]   A method for partial volume correction of PET-imaged tumor heterogeneity using expectation maximization with a spatially varying point spread function [J].
Barbee, David L. ;
Flynn, Ryan T. ;
Holden, James E. ;
Nickles, Robert J. ;
Jeraj, Robert .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01) :221-236
[7]  
Bardiès M, 2011, Q J NUCL MED MOL IM, V55, P5
[8]   NOISE PROPERTIES OF THE EM ALGORITHM .1. THEORY [J].
BARRETT, HH ;
WILSON, DW ;
TSUI, BMW .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (05) :833-846
[9]   Brain PET partial-volume compensation using blurred anatomical labels [J].
Bataille, F. ;
Comtat, C. ;
Jan, S. ;
Sureau, F. C. ;
Trebossen, R. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (05) :1606-1615
[10]   Study of relative quantification of Tc-99 m with partial volume effect and spillover correction for SPECT oncology imaging [J].
Boening, Guido ;
Pretorius, P. Hendrik ;
King, Michael A. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1205-1212