Estimation of beat-to-beat changes in stroke volume from arterial pressure: a comparison of two pressure wave analysis techniques during head-up tilt testing in young, healthy men

被引:42
作者
Jellema, WT
Imholz, BPM
Oosting, H
Wesseling, KH
van Lieshout, JJ
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Internal Med, NL-1100 DE Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Clin Epidemiol & Biostat, NL-1100 DE Amsterdam, Netherlands
[3] Univ Amsterdam, Acad Med Ctr, TNO BioMed Instrumentat, NL-1100 DE Amsterdam, Netherlands
关键词
aortic impedance; blood pressure; Finapres; head-up tilt; pulse contour; simulation; stroke volume;
D O I
10.1007/BF02330482
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: The aim of this study was to compare beat-to-beat changes in stroke volume (SV) estimated by two different pressure wave analysis techniques during orthostatic stress testing: pulse contour analysis and Modelflow, ie, simulation of a three-element model of aortic input impedance. Methods: A reduction in SV was introduced in eight healthy young men (mean age, 25; range, 19-32 y) by a 30-minute head-up tilt maneuver. Intrabrachial and noninvasive finger pressure were monitored simultaneously. Beat-to-beat changes in SV were estimated from intrabrachial pressure by pulse contour analysis and Modelflow. In addition, the relative differences in Modelflow SV obtained from intrabrachial pressure and noninvasive finger pressure were assessed. Results: Beat-to-beat changes in Modelflow SV from intrabrachial pressure were comparable with pulse contour measures. The relative difference between the two methods amounted to 0.1 +/- 1% (mean +/- SEM) and was not dependent on the duration of tilt. The difference between Modelflow applied to intrabrachial pressure and finger pressure amounted to -2.7 +/- 1.3% (p = 0.04). This difference was not dependent on the duration of tilt or level of arterial pressure. Conclusions: Based on different mathematical models of the human arterial system, pulse contour and Modelflow compute similar changes in SV from intrabrachial pressure during orthostatic stress testing in young healthy men. The magnitude of the difference in SV derived from intrabrachial and finger pressure may vary among subjects; Modelflow SV from noninvasive finger pressure tracks fast and brisk changes in SV derived from intrabrachial pressure.
引用
收藏
页码:185 / 192
页数:8
相关论文
共 33 条
[1]   Tilt table testing for assessing syncope [J].
Benditt, DG ;
Ferguson, DW ;
Grubb, BP ;
Kapoor, WN ;
Kugler, J ;
Lerman, BB ;
Maloney, JD ;
Raviele, A ;
Ross, B ;
Sutton, R ;
Wolk, MJ ;
Wood, DL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (01) :263-275
[2]   THE RELIABILITY OF NONINVASIVE CONTINUOUS FINGER BLOOD-PRESSURE MEASUREMENT IN PATIENTS WITH BOTH HYPERTENSION AND VASCULAR-DISEASE [J].
BOS, WJW ;
IMHOLZ, BPM ;
VANGOUDOEVER, J ;
WESSELING, KH ;
VANMONTFRANS, GA .
AMERICAN JOURNAL OF HYPERTENSION, 1992, 5 (08) :529-535
[3]   Reconstruction of brachial artery pressure from noninvasive finger pressure measurements [J].
Bos, WJW ;
vanGoudoever, J ;
vanMontfrans, GA ;
vandenMeiracker, AH ;
Wesseling, KH .
CIRCULATION, 1996, 94 (08) :1870-1875
[4]  
BURKHOFF A, 1988, AM J PHYSIOL, V255, pH742
[5]   IMPROVED METHOD FOR CARDIAC-OUTPUT DETERMINATION IN MAN USING ULTRASOUND DOPPLER TECHNIQUE [J].
ERIKSEN, M ;
WALLOE, L .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (06) :555-560
[6]   COMPARISON OF CARDIAC-OUTPUT MEASUREMENT TECHNIQUES - THERMODILUTION, DOPPLER, CO2-REBREATHING AND THE DIRECT FICK METHOD [J].
ESPERSEN, K ;
JENSEN, EW ;
ROSENBORG, D ;
THOMSEN, JK ;
ELIASEN, K ;
OLSEN, NV ;
KANSTRUP, IL .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 (02) :245-251
[7]  
HARMS MPM, 1999, IN PRESS CLIN SCI
[8]   Noninvasive assessment of cardiac output in critically ill patients by analysis of the finger blood pressure waveform [J].
Hirschl, MM ;
Binder, M ;
Gwechenberger, M ;
Herkner, H ;
Bur, A ;
Kittler, H ;
Laggner, AN .
CRITICAL CARE MEDICINE, 1997, 25 (11) :1909-1914
[9]   Fifteen years experience with finger arterial pressure monitoring: assessment of the technology [J].
Imholz, BPM ;
Wieling, W ;
van Montfrans, GA ;
Wesseling, KH .
CARDIOVASCULAR RESEARCH, 1998, 38 (03) :605-616
[10]   CONTINUOUS CARDIAC-OUTPUT MONITORING BY PULSE CONTOUR DURING CARDIAC-SURGERY [J].
JANSEN, JRC ;
WESSELING, KH ;
SETTELS, JJ ;
SCHREUDER, JJ .
EUROPEAN HEART JOURNAL, 1990, 11 :26-32