Computed tomography-based navigation for hip, knee, and spine surgery

被引:63
作者
Amiot, LP
Poulin, F
机构
[1] Univ Montreal, Hop Maison Neuve Rosemont, Dept Orthoped, Montreal, PQ H1T 4B3, Canada
[2] ORTHOsoft Inc, Montreal, PQ, Canada
关键词
D O I
10.1097/01.blo.0000126866.29933.42
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A review of CT-based orthopaedic navigation is presented with a specific emphasis on arthroplasty for the hip and the knee. Fundamental issues about the laboratory and clinical validation of the applications are addressed. The ability to compute the position and orientation of an acetabular implant using a postoperative CT scan was investigated. Angle deviations relative to known positions were computed with an error of less than 1 degree. Then, the system accuracy for three-dimensional reconstruction and registration of two cadaveric pelvis specimens was measured with more than 350 registrations. We observed a maximal inclination error of 5degrees in 99% of cases and a maximal anteversion error of 5degrees in 97% of cases. The accuracy of the three-dimensional reconstruction and registration for knee arthroplasty also was measured and computed with an angular accuracy of 0.5degrees in the AP plane and accuracy of 3degrees in the lateral plane. A clinical study then was done in 109 cases where 96% of implants were installed with a hip-knee-ankle angle of 180 +/- 3degrees. Computed tomography-based navigation for orthopaedic surgery provides greater accuracy and reproducibility than conventional surgery. As noted by learning curves, software improvements are needed to bring it into daily clinical routine.
引用
收藏
页码:77 / 86
页数:10
相关论文
共 62 条
[1]   COMPUTER-ASSISTED PEDICLE SCREW FIXATION - A FEASIBILITY STUDY [J].
AMIOT, LP ;
LABELLE, H ;
DEGUISE, JA ;
SATI, M ;
BRODEUR, P ;
RIVARD, CH .
SPINE, 1995, 20 (10) :1208-1212
[2]   Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine [J].
Amiot, LP ;
Lang, K ;
Putzier, M ;
Zippel, H ;
Labelle, H .
SPINE, 2000, 25 (05) :606-614
[3]  
Amiot LP, 1996, ANN CHIR, V50, P626
[4]   Transpedicular screw placement -: Image-guided versus lateral-view fluoroscopy:: In vitro simulation [J].
Assaker, R ;
Reyns, N ;
Vinchon, M ;
Demondion, X ;
Louis, E .
SPINE, 2001, 26 (19) :2160-2164
[5]  
Bachler R, 2001, Comput Aided Surg, V6, P143
[6]  
Barrick E F, 1998, Comput Aided Surg, V3, P289, DOI 10.3109/10929089809148149
[7]  
Bernsmann K, 2001, Z ORTHOP GRENZGEB, V139, P512
[8]  
Borner M, 1999, Z Arztl Fortbild Qualitatssich, V93, P253
[9]   Factors influencing accuracy of screw displacement axis detection with a DC-based electromagnetic tracking system [J].
Bottlang, M ;
Marsh, JL ;
Brown, TD .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1998, 120 (03) :431-435
[10]  
Butler W E, 1999, Comput Aided Surg, V4, P26