A comparison of the imaging properties of CCD-based devices used for small field digital mammography

被引:52
作者
Evans, DS
Workman, A
Payne, M
机构
[1] Kings Coll London Hosp, KCARE, London SE5 9RS, England
[2] Forster Green Hosp, NIRMPA, Belfast BT8 4DE, Antrim, North Ireland
关键词
D O I
10.1088/0031-9155/47/1/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective imaging characteristics of three systems that use charge coupled devices (CCD) for small-field digital mammography (SFDM) have been compared in terms of spatial resolution and signal to noise ratio. The results indicate that although they are designed for nominally the same tasks of stereotactic localization and spot imaging these detectors have significantly differing physical imaging properties. Imaging system design parameters such as the phosphor screen type and thickness, screen configuration and method of optically coupling the phosphor to the CCD have significant effects on the imaging performance of the detectors.
引用
收藏
页码:117 / 135
页数:19
相关论文
共 42 条
[1]   Spectral modeling and compilation of quantum fluence in radiography and mammography [J].
Boone, JM .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :592-601
[2]   A Monte Carlo study of x-ray fluorescence in x-ray detectors [J].
Boone, JM ;
Seibert, JA ;
Sabol, JM ;
Tecotzky, M .
MEDICAL PHYSICS, 1999, 26 (06) :905-916
[3]   A CLINICAL COMPARISON BETWEEN CONVENTIONAL AND DIGITAL MAMMOGRAPHY UTILIZING COMPUTED RADIOGRAPHY [J].
BRETTLE, DS ;
WARD, SC ;
PARKIN, GJS ;
COWEN, AR ;
SUMSION, HJ .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (797) :464-468
[4]   Advances in high-speed mammographic image quality [J].
Bunch, PC .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :120-130
[5]   ANALYSIS OF THE DETECTIVE QUANTUM EFFICIENCY OF A RADIOGRAPHIC SCREEN FILM COMBINATION [J].
BUNCH, PC ;
HUFF, KE ;
VANMETTER, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :902-909
[6]  
Cranley K., 1997, CATALOGUE DIAGNOSTIC
[7]   A METHOD FOR MODULATION TRANSFER-FUNCTION DETERMINATION FROM EDGE PROFILES WITH CORRECTION FOR FINITE-ELEMENT DIFFERENTIATION [J].
CUNNINGHAM, IA ;
FENSTER, A .
MEDICAL PHYSICS, 1987, 14 (04) :533-537
[8]  
Dainty JC., 1974, IMAGE SCI
[9]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[10]   EFFECTS OF UNDERSAMPLING ON THE PROPER INTERPRETATION OF MODULATION TRANSFER-FUNCTION, NOISE POWER SPECTRA, AND NOISE EQUIVALENT QUANTA OF DIGITAL IMAGING-SYSTEMS [J].
DOBBINS, JT .
MEDICAL PHYSICS, 1995, 22 (02) :171-181