A comparison of the imaging properties of CCD-based devices used for small field digital mammography

被引:52
作者
Evans, DS
Workman, A
Payne, M
机构
[1] Kings Coll London Hosp, KCARE, London SE5 9RS, England
[2] Forster Green Hosp, NIRMPA, Belfast BT8 4DE, Antrim, North Ireland
关键词
D O I
10.1088/0031-9155/47/1/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective imaging characteristics of three systems that use charge coupled devices (CCD) for small-field digital mammography (SFDM) have been compared in terms of spatial resolution and signal to noise ratio. The results indicate that although they are designed for nominally the same tasks of stereotactic localization and spot imaging these detectors have significantly differing physical imaging properties. Imaging system design parameters such as the phosphor screen type and thickness, screen configuration and method of optically coupling the phosphor to the CCD have significant effects on the imaging performance of the detectors.
引用
收藏
页码:117 / 135
页数:19
相关论文
共 42 条
[11]   Modulation transfer function for a large-area amorphous silicon image receptor [J].
Earnhart, JRD ;
Chaney, EL .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (12) :2531-2540
[12]  
Fujita H, 1989, P SOC PHOTO-OPT INS, V1090, P263
[13]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .2. NOISE WIENER SPECTRUM [J].
GIGER, ML ;
DOI, K ;
METZ, CE .
MEDICAL PHYSICS, 1984, 11 (06) :797-805
[14]   IMAGE-INFORMATION TRANSFER PROPERTIES OF X-RAY INTENSIFYING SCREENS IN THE ENERGY-RANGE FROM 17 TO 320 KEV [J].
GINZBURG, A ;
DICK, CE .
MEDICAL PHYSICS, 1993, 20 (04) :1013-1021
[15]   Performance of a 41X41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL PHYSICS, 2000, 27 (06) :1324-1331
[16]   System considerations in CCD-based x-ray imaging for digital chest radiography and digital mammography [J].
Hejazi, S ;
Trauernicht, DP .
MEDICAL PHYSICS, 1997, 24 (02) :287-297
[17]  
JARLMAN O, 1991, ACTA RADIOL, V32, P110
[18]   Detective quantum efficiency of a CsI:Tl scintillator based scanning slot x-ray detector for digital mammography [J].
Jing, ZX ;
Huda, W ;
Walker, JK ;
Choi, WY .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :583-591
[19]   Digital mammography performed with computed radiography technology [J].
Jouan, B .
EUROPEAN JOURNAL OF RADIOLOGY, 1999, 31 (01) :18-24
[20]   CHARGE-COUPLED DEVICE DETECTOR - PERFORMANCE CONSIDERATIONS AND POTENTIAL FOR SMALL-FIELD MAMMOGRAPHIC IMAGING APPLICATIONS [J].
KARELLAS, A ;
HARRIS, LJ ;
LIU, H ;
DAVIS, MA ;
DORSI, CJ .
MEDICAL PHYSICS, 1992, 19 (04) :1015-1023