A QCM immunosensor for Salmonella detection with simultaneous measurements of resonant frequency and motional resistance

被引:140
作者
Su, XL [1 ]
Li, YB [1 ]
机构
[1] Univ Arkansas, Dept Biol & Agr Engn, Fayetteville, AR 72701 USA
关键词
quartz crystal microbalance; immunosensor; Salmonella Typhimurium; resonant frequency; motional resistance;
D O I
10.1016/j.bios.2005.01.021
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A quartz crystal microbalance (QCM) immunosensor was described for the detection of Salmonella Typhimurium with simultaneous measurements of the resonant frequency and motional resistance. The immunosensor was fabricated using protein A for the antibody immobilization. High-frequency impedance analysis indicated that the changes in resonant frequency and motional resistance (Delta F and Delta R) of the QCM were significant while the changes in static capacitance, motional capacitance, and motional inductance were insignificant. In the direct detection of S. Typhimurium in chicken meat sample, Delta F and Delta R were proportional to the cell concentration in the range of 10(5)-10(8) and 10(6)-10(8) cells/ml, respectively. Using anti-Salmonella-magnetic beads as a separator/concentrator for sample pretreatment as well as a marker for signal amplification, the detection limit was lowered to 10(2) cells/ml based on the Delta R measurement; however, Delta F was not related to the cell concentration. No interference was observed from E. coli K12 or the sample matrix. 0 2005 Elsevier B.V. All fights reserved.
引用
收藏
页码:840 / 848
页数:9