Primary stability of four different implants for opening wedge high tibial osteotomy

被引:181
作者
Agneskirchner, JD
Freiling, D
Hurschler, C
Lobenhoffer, P
机构
[1] Henriettenstiftung Hannover, Dept Trauma & Reconstruct Surg, D-30171 Hannover, Germany
[2] Hannover Med Sch, Dept Orthopaed, Lab Biomech & Biomat, D-3000 Hannover, Germany
关键词
osteoarthritis; high tibial osteotomy; fixation method; Puddu plate; Tomofix; locking compression plate;
D O I
10.1007/s00167-005-0690-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Since a significant number of implant failures have been reported in association with the procedure of open wedge valgus high tibial osteotomy, the initial biomechanical stability of different fixation devices was investigated in this study. Fifteen third generation Sawbones composite tibiae were used as a model. Four different plates were tested: a short spacer plate (OWO) (n=4), a short spacer plate with multi-directional locking bolts (MSO) (n=5), a prototype version of a long spacer plate with multi-directional locking bolts (MSOnew) (n=2), and a long medial tibia plate fixator with locking bolts (MPF) (n=4). All opening wedge osteotomies were performed by the same surgeon (PL) in a standardized fashion. Axial compression of the tibiae was performed using a materials testing machine under standardized alignment of the loading axis. Single load to failure tests as well as load-controlled cyclical failure tests were performed. The required force and cycles to failure were recorded. Osteotomy gap motion was measured using linear displacement transducers. Residual stability after failure of the opposite lateral cortex was analysed. Failure occurred at the lateral cortex bone-bride in all tested implants. The rigid long plate fixator (MPF) resisted the greatest amount of force (2,881 N) in the single load to failure tests. In the cyclical load-to-failure tests, the constructs with MPF resisted more than twice the amount of loading cycles when compared to the short spacer plates. The osteotomy gap motion was smallest in the MPF, with a reduction of the displacements of up to 65, 66 and 88%, when compared to OWO, MSO and MSOnew, respectively. The highest residual stability after failure of the lateral cortex was observed in MPF as well. The results suggest that the implant design strongly influences the primary stability of medial opening wedge tibial osteotomy. A rigid long plate fixator with angle-stable locking bolts yields the best results.
引用
收藏
页码:291 / 300
页数:10
相关论文
共 15 条
[1]   Drop foot after high tibial osteotomy: a prospective study of aetiological factors [J].
Bauer, T ;
Hardy, P ;
Lemoine, J ;
Finlayson, DF ;
Tranier, S ;
Lortat-Jacob, A .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2005, 13 (01) :23-33
[2]   PROXIMAL TIBIAL OSTEOTOMY - A CRITICAL LONG-TERM STUDY OF 87 CASES [J].
COVENTRY, MB ;
ILSTRUP, DM ;
WALLRICHS, SL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (02) :196-201
[3]   Mechanical validation of whole bone composite tibia models [J].
Cristofolini, L ;
Viceconti, M .
JOURNAL OF BIOMECHANICS, 2000, 33 (03) :279-288
[4]   Primary stability of different implants used in conjunction with high tibial osteotomy [J].
Flamme, CH ;
Kohn, D ;
Kirsch, L ;
Hurschler, C .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1999, 119 (7-8) :450-455
[5]  
FUJISAWA Y, 1979, ORTHOP CLIN N AM, V10, P585
[6]   Structural properties of a new design of composite replicate femurs and tibias [J].
Heiner, AD ;
Brown, TD .
JOURNAL OF BIOMECHANICS, 2001, 34 (06) :773-781
[7]  
Korovessis P, 1999, ORTHOPEDICS, V22, P729
[8]   Open-wedge high tibial osteotomy with special medial plate fixator [J].
Lobenhoffer P. ;
Agneskirchner J. ;
Zoch W. .
Der Orthopäde, 2004, 33 (2) :153-160
[9]   Improvements in surgical technique of valgus high tibial osteotomy [J].
Lobenhoffer, P ;
Agneskirchner, JD .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2003, 11 (03) :132-138
[10]  
Lobenhoffer P., 2002, TECH KNEE SURG, V1, P153, DOI [10.1097/00132588-200212000-00004, DOI 10.1097/00132588-200212000-00004]