Type of solution and PCO2 measurement errors during tonometry

被引:20
作者
Kolkman, JJ
Zwarekant, LJ
Boshuizen, K
Groeneveld, ABJ
Steverink, PJGM
Meuwissen, SGM
机构
[1] FREE UNIV AMSTERDAM HOSP,DEPT GASTROENTEROL,AMSTERDAM,NETHERLANDS
[2] FREE UNIV AMSTERDAM HOSP,DEPT CLIN PHYS & ENGN,AMSTERDAM,NETHERLANDS
[3] FREE UNIV AMSTERDAM HOSP,MED INTENS CARE UNIT,AMSTERDAM,NETHERLANDS
[4] UNIV UTRECHT,DEPT GEN & LARGE ANIM SURG,UTRECHT,NETHERLANDS
关键词
PCO2; tonometry; measurement errors; intramucosal pH; blood gas analyzers; carbon dioxide;
D O I
10.1007/s001340050390
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: The choice of solution for gastrointestinal tonometry influences the PCO2 measurement bias, precision and the time required for equilibration. We compared saline with buffered solutions during in vitro tonometry, with respect to systematic and accidental measurement errors and equilibration time. Design: A prospective laboratory study. Measurements: Saline, phosphate. phosphate bicarbonate and succinylated gelatin solutions were equilibrated in a specialized blood gas tonometer at PCO(2)s of 2.7, 3.6, 4.5, 6.2 and 9.0 kPa, using calibration gases. Accidental errors were determined: the within-syringe decline of PCO2 and the effects of handling errors (five up and down movements of the plunger). The PCO2 build up in gastrointestinal tonometers was determined in 5000 ml saline baths with fixed PCO2 levels of 2.7 and 9.0 kPa. Results: The build up of PCO2 in phosphate bicarbonate and gelatin was about 4 and 2 times slower than in saline and phosphate, respectively, both for gas and gastrointestinal tonometers. The bias of the measured PCO2 at equilibrium was -15 % for saline, and between -1 and 3 % for phosphate. phosphate bicarbonate and gelatin. The precision was comparable among the solutions: 2 +/- 1% for saline, 2 +/- 1 % for phosphate. 1 +/- 0 % for phosphate bicarbonate and 1 +/- 1 % for gelatin. The accidental errors were virtually absent with phosphate bicarbonate, intermediate with gelatin and largest with saline and phosphate. Conclusion: Phosphate bicarbonate buffer and succinylated gelatin allow accurate PCO2 measurements, but their equilibration is too slow for clinical application. The advantage of phosphate over saline solution is a smaller bias only. Thus, both saline and phosphate are currently the tonometer solutions of choice, provided that strictly anaerobic conditions are applied and the bias by the blood gas analyzer is known.
引用
收藏
页码:658 / 663
页数:6
相关论文
共 8 条
[1]  
Crispin C, 1995, Intensive Crit Care Nurs, V11, P123, DOI 10.1016/S0964-3397(95)80592-3
[2]  
FIDDIANGREEN RG, 1986, ARCH SURG-CHICAGO, V121, P654
[3]   IMPROVED PCO2 MEASUREMENT IN 6 STANDARD BLOOD-GAS ANALYZERS USING A PHOSPHATE-BUFFERED SOLUTION FOR GASTRIC TONOMETRY [J].
KNICHWITZ, G ;
MERTES, N ;
KUHMANN, M .
ANAESTHESIA, 1995, 50 (06) :532-534
[4]   Gastric tonometry: Precision and reliability are improved by a phosphate buffered solution [J].
Knichwitz, G ;
Kuhmann, M ;
Brodner, G ;
Mertes, N ;
Goeters, C ;
Brussel, T .
CRITICAL CARE MEDICINE, 1996, 24 (03) :512-516
[5]   MEASURING CARBON-DIOXIDE TENSION IN SALINE AND ALTERNATIVE SOLUTIONS - QUANTIFICATION OF BIAS AND PRECISION IN 2 BLOOD-GAS ANALYZERS [J].
RIDDINGTON, D ;
VENKATESH, KB ;
CLUTTONBROCK, T ;
BION, J .
CRITICAL CARE MEDICINE, 1994, 22 (01) :96-100
[6]  
TAKALA J, 1994, CRIT CARE MED, V22, P1877
[7]  
VEEFKIND AH, 1975, CLIN CHEM, V21, P685
[8]   POTENTIAL HAZARDS IN ESTIMATION OF GASTRIC INTRAMUCOSAL PH [J].
WOOD, PR ;
CUDWORTH, P ;
DAVIES, GE ;
LAWLER, PG .
LANCET, 1992, 340 (8824) :913-913