Hemodynamic performance of the PRIMA(TM) Edwards stentless aortic xenograft: Early results of a multicenter clinical trial

被引:17
作者
Dossche, K
Vanermen, H
Daenen, W
Pillai, R
Konertz, W
机构
[1] CATHOLIC UNIV LEUVEN,B-3000 LOUVAIN,BELGIUM
[2] OXFORD HEART CTR,OXFORD,ENGLAND
[3] UNIV HOSP CHARITE,W-1000 BERLIN,GERMANY
关键词
stentless aortic valve; xenograft; hemodynamics; early results;
D O I
10.1055/s-2007-1011975
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Between lune 1991 and July 1993, 200 PRIMA(TM) Edwards stentless aortic bioprostheses were implanted in 4 European centres. Mean age of the group was 68.5 years (S.D. 8.0). The subcoronary implantation technique was used in 161 patients, the inclusion cylinder technique in 39 patients. Valve diameters ranged from 19 mm to 29 mm. Transthoracic Doppler echocardiographic studies were performed at discharge, and after 6 months and 12 months; a complete one-year follow-up was obtained. At 12 months peak systolic gradients ranged from 35.2 mmHg (19 mm valve) to 10.9 mmHg (29 mm valve) and effective orifice area ranged from 0.8 cm(2) (19 mm valve) to 2.8 cm(2) (29 mm valve). 30-day mortality was 3%, 12-month mortality was 5%. Complications were thrombo-embolism in six patients (3%), haemorrhage in three patients (1.5%), endocarditis in two patients (1%), and total AV-block requiring an endocavitary pacemaker in 14 patients (7%). At one year aortic regurgitation was evident in 27% patients but only one patient showed grade III. The stentless xenograft still offers good hemodynamics at one year. Morbidity and mortality are acceptable. Evaluation of the long-term performance of this new type of xenograft will be an important issue in future.
引用
收藏
页码:11 / 14
页数:4
相关论文
共 16 条
[1]   DEATH AND OTHER TIME-RELATED EVENTS AFTER VALVE-REPLACEMENT [J].
BLACKSTONE, EH ;
KIRKLIN, JW .
CIRCULATION, 1985, 72 (04) :753-767
[2]   BIOLOGICAL FACTORS AFFECTING LONG-TERM RESULTS OF VALVULAR HETEROGRAFTS [J].
CARPENTIER, A ;
LEMAIGRE, G ;
ROBERT, L ;
CARPENTIER, S ;
DUBOST, C ;
GERBODE, F .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1969, 58 (04) :467-+
[3]   CONTINUOUS-WAVE DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF SEVERITY OF CALCIFIC AORTIC-STENOSIS - A SIMULTANEOUS DOPPLER-CATHETER CORRELATIVE STUDY IN 100 ADULT PATIENTS [J].
CURRIE, PJ ;
SEWARD, JB ;
REEDER, GS ;
VLIETSTRA, RE ;
BRESNAHAN, DR ;
BRESNAHAN, JF ;
SMITH, HC ;
HAGLER, DJ ;
TAJIK, AJ .
CIRCULATION, 1985, 71 (06) :1162-1169
[4]  
DAVID TE, 1990, J THORAC CARDIOV SUR, V99, P113
[5]  
DURY PJ, 1986, BIOL BIOPROSTHESIS V, P580
[6]   THE VALUE OF THE STENTLESS BIOVALVE PROSTHESIS - AN EXPERIMENTAL-STUDY [J].
HAZEKAMP, MG ;
GOFFIN, YA ;
HUYSMANS, HA ;
CICHON, A .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1993, 7 (10) :514-519
[7]   TECHNIQUE OF AORTIC-VALVE REPLACEMENT WITH THE EDWARDS STENTLESS AORTIC BIOPROSTHESIS 2500 [J].
KONERTZ, W ;
WEYAND, M ;
SIDIROPOULOS, A ;
SCHWAMMENTHAL, E ;
BREITHARDT, G ;
SCHELD, HH .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (05) :274-277
[8]   LONG-TERM PERFORMANCE OF 555 AORTIC HOMOGRAFTS IN THE AORTIC POSITION [J].
MATSUKI, O ;
ROBLES, A ;
GIBBS, S ;
BODNAR, E ;
ROSS, DN .
ANNALS OF THORACIC SURGERY, 1988, 46 (02) :187-191
[9]  
O'Brien M F, 1993, J Heart Valve Dis, V2, P385
[10]   STENTLESS AORTIC BIOPROSTHESIS - THE WAY FORWARD - EARLY EXPERIENCE WITH THE EDWARDS VALVE [J].
PILLAI, R ;
SPRIGGINGS, D ;
AMARASENA, N ;
OREGAN, DJ ;
PARRY, AJ ;
WESTABY, S .
ANNALS OF THORACIC SURGERY, 1993, 56 (01) :88-91