On-line monitoring of chromium aerosols in industrial exhaust streams by laser-induced plasma spectroscopy (LIPS)

被引:33
作者
Neuhauser, RE
Panne, U
Niessner, R
Wilbring, P
机构
[1] Tech Univ Munich, Inst Hydrochem, D-81377 Munich, Germany
[2] Inst Environm Protect & Energy Technol, Air Pollut Control Dept, D-51105 Cologne, Germany
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1999年 / 364卷 / 08期
关键词
D O I
10.1007/s002160051421
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ultrafine heavy metal aerosols can have severe toxic and carcinogenic effects when inhaled in higher concentrations. The objective of this work was to develop a mobile continuous emission monitor (CEM) based on laser-induced plasma spectroscopy (LIPS) for an on-line analysis of chromium aerosols. The LIPS emission sensor was installed in an electroplating facility for a fast monitoring of particulate emissions to provide instantaneous feedback for process control. The prototype was tested in cooperation with an independent laboratory, which determined the total chromium content and the Cr(VI)/Cr(III) ratio in the exhaust stream by conventional filter analysis. The system provided both the necessary time-resolution and detection limits (14 mu g m(-3)) for emission monitoring below the legal threshold (1 mg m(-3)) values. A good correlation was found between the on-line LIPS measurement and the independent reference analysis.
引用
收藏
页码:720 / 726
页数:7
相关论文
共 47 条