Quality control of CD4+ T-lymphocyte enumeration:: Results from the last 9 years of the United Kingdom National External Quality Assessment Scheme for Immune Monitoring (1993-2001)

被引:71
作者
Whitby, L [1 ]
Granger, V [1 ]
Storie, I [1 ]
Goodfellow, K [1 ]
Sawle, A [1 ]
Reilly, JT [1 ]
Barnett, D [1 ]
机构
[1] Royal Hallamshire Hosp, Dept Haematol, UK NEQAS Leucocyte Immunophenotyping, Sheffield, S Yorkshire, England
来源
CYTOMETRY | 2002年 / 50卷 / 02期
关键词
flow cytometry; CD4(+) T lymphocytes; quality control; quality assessment;
D O I
10.1002/cyto.10094
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The human immunodeficiency virus (HIV) global epidemic has necessitated the routine enumeration of T-lymphocyte subsets, which has created a need for external quality assurance (EQA). The United Kingdom National External Quality Assessment Scheme (UK NEQAS) for Immune Monitoring provides EQA for 296 laboratories in 40 countries. In 1993, UK NEQAS developed and incorporated into its program stabilized whole blood that enables the accurate monitoring of laboratory performance. Overall, the mean interlaboratory coefficient of variation (CV) for percentage CD4(+) T-lymphocyte subset enumeration has fallen from 15% to less than 5%, as a direct result of the increased use of CD45/ side scatter (SSC) gating. Laboratories using alternative gating strategies (i.e., CD45/CD14 or forward scatter [FSC]/SSC) were about 7.4 times more likely to fail an EQA exercise. Furthermore, the adoption of single-platform technology resulted in a reduction of the overall mean interlaboratory CV for absolute CD4+ T lymphocytes from 56% (prior to the widespread use of single-platform technology) to 9.7%. Individual laboratory deficiencies were also identified using a performance monitoring system and, through re-education by collaboration with the coordinating center, satisfactorily resolved. In conclusion, during the last 9 years, the UK NEQAS for Immune Monitoring program has highlighted the significant technological advances made by laboratories worldwide that undertake lymphocyte subset enumeration. Cytometry (Clin. Cytometry) 50:102-110, 2002. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:102 / 110
页数:9
相关论文
共 30 条
[1]  
Barnett D, 1996, CYTOMETRY, V26, P216, DOI 10.1002/(SICI)1097-0320(19960915)26:3<216::AID-CYTO6>3.0.CO
[2]  
2-C
[3]   Reduction of intra- and interlaboratory variation in CD34+ stem cell enumeration using stable test material, standard protocols and targeted training [J].
Barnett, D ;
Granger, V ;
Kraan, J ;
Whitby, L ;
Reilly, JT ;
Papa, S ;
Gratama, JW .
BRITISH JOURNAL OF HAEMATOLOGY, 2000, 108 (04) :784-792
[4]  
Barnett D, 1999, BRIT J HAEMATOL, V106, P1059
[5]   Guidelines for the enumeration of CD4+ T lymphocytes in immunosuppressed individuals [J].
Barnett, D ;
Bird, G ;
Hodges, E ;
Linch, DC ;
Matutes, E ;
Newland, AC ;
Reilly, JT ;
Bain, BJ ;
Amos, R ;
Chapman, C ;
England, JM ;
Hyde, K .
CLINICAL AND LABORATORY HAEMATOLOGY, 1997, 19 (04) :231-241
[6]  
BARNETT D, 1998, Patent No. 2279653
[7]  
Bergeron M, 1998, CYTOMETRY, V33, P146, DOI 10.1002/(SICI)1097-0320(19981001)33:2<146::AID-CYTO9>3.0.CO
[8]  
2-H
[9]   PREDICTIVE VALUE OF REPEATED MEASUREMENTS OF CD4 LYMPHOCYTE COUNTS ON PROGRESSION TO AIDS [J].
BOUTITIE, F ;
POCOCK, SJ .
AIDS, 1994, 8 (01) :35-41
[10]  
Centers for Disease Control, 1993, MMWR Morb Mortal Wkly Rep, V41, P1