Percutaneous aortic valve replacement: An experimental study. I. Studies on implantation

被引:103
作者
Lutter, G
Kuklinski, D
Berg, G
von Samson, P
Martin, J
Handke, M
Uhrmeister, P
Beyersdorf, F
机构
[1] Univ Freiburg, Med Ctr, Div Cardiovasc Surg, Sch Med, D-79106 Freiburg, Germany
[2] Univ Freiburg, Sch Med, Div Cardiol, D-79106 Freiburg, Germany
[3] Univ Freiburg, Sch Med, Div Radiol, D-79106 Freiburg, Germany
关键词
D O I
10.1067/mtc.2002.121157
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The purpose of this preliminary study was to devise a new surgical procedure for minimally invasive aortic valve implantation with a transluminal technique. Methods: The new collapsible heart valve was prepared by mounting a porcine aortic valve, taken from a freshly slaughtered pig, into a self-expandable nitinol stent by means of a suture technique. The outer diameter of the valved stent ranged from 15 to 23 mm, and the length ranged from 21 to 28 mm. Before implantation in vivo, these valved stents were tested in an in vitro circulatory system. Only in vitro-tested valved stents with a pressure Gradient of less than 7 mm Hg and regurgitation of 1degrees or less were used for transluminal aortic valve implantation in vivo. Six of these valved stents were implanted in the descending aorta and 8 in the ascending aorta of anesthetized pigs. The catheter delivery system (22F) was extraperitoneally inserted through the left iliac artery or the infrarenal aorta. Measurements for transvalvular gradient, valvular opening and closure, blood-flow characteristics, regurgitation, and macroscopic analysis were performed at baseline and after the observation period (164 +/- 48 minutes). Results: This preliminary study contained 14 animals. One animal died of ventricular fibrillation. Technical failure occurred in 2 pigs as a result of stent twisting. At the end of the observation period, the 11 successfully implanted valved stents demonstrated low transvalvular gradients (mean end-systolic Deltarho(max) of 5.4 +/- 3.3 mm Hg for the descending aorta group, 5.4 +/- 1.2 mm Hg for the supracoronary group, and 5.4 +/- 1.1 mm Hg for the subcoronary group), which did not differ from their in vitro gradients. Two-dimensional echocardiography demonstrated complete valvular closure and opening in 5 of 5 cases. Angiography indicated only a physiologic jet of regurgitation (0degrees) in 8 animals and mild (1degrees) regurgitation in 3 animals. Color Doppler ultrasonography indicated no regurgitation in 5 of 5 cases and minor paravalvular leakage in 1 case. Conclusion: Aortic valved stents can be successfully implanted without thoracotomy by using a transluminal catheter technique. Lone-term function of the valves remains to be established.
引用
收藏
页码:768 / 776
页数:9
相关论文
共 12 条
[1]   TRANSLUMINAL IMPLANTATION OF ARTIFICIAL-HEART VALVES - DESCRIPTION OF A NEW EXPANDABLE AORTIC-VALVE AND INITIAL RESULTS WITH IMPLANTATION BY CATHETER TECHNIQUE IN CLOSED CHEST PIGS [J].
ANDERSEN, HR ;
KNUDSEN, LL ;
HASENKAM, JM .
EUROPEAN HEART JOURNAL, 1992, 13 (05) :704-708
[2]   Transcatheter implantation of a bovine valve in pulmonary position - A lamb study [J].
Bonhoeffer, P ;
Boudjemline, Y ;
Saliba, Z ;
Hausse, AO ;
Aggoun, Y ;
Bonnet, D ;
Sidi, D ;
Kachaner, J .
CIRCULATION, 2000, 102 (07) :813-816
[3]   The "first generation" of endovascular stent-grafts for patients with aneurysms of the descending thoracic aorta [J].
Dake, MD ;
Miller, DC ;
Mitchell, RS ;
Semba, CP ;
Moore, KA ;
Sakai, T .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1998, 116 (05) :689-703
[4]  
HILBERT SL, 1987, J THORAC CARDIOV SUR, V94, P419
[5]   Transmyocardial laser revascularization: Experimental studies in healthy porcine myocardium [J].
Lutter, G ;
Martin, J ;
Takahashi, N ;
Yoshitake, M ;
Schwarzkopf, J ;
Nitzsche, E ;
Beyersdorf, F .
ANNALS OF THORACIC SURGERY, 1999, 67 (06) :1708-1713
[6]  
LUTTER G, 2001, THORAC CARIODVASC S1, V49, P67
[7]  
MAOZAMI N, 1996, ASAIO J, V42, pM381
[8]   The stent decade: 1987 to 1997 [J].
Oesterle, SN ;
Whitbourn, R ;
Fitzgerald, PJ ;
Yeung, AC ;
Stertzer, SH ;
Dake, MD ;
Yock, PG ;
Virmani, R .
AMERICAN HEART JOURNAL, 1998, 136 (04) :578-599
[9]   DEVELOPMENT AND INITIAL EXPERIMENTAL EVALUATION OF A PROSTHETIC AORTIC-VALVE FOR TRANSCATHETER PLACEMENT - WORK IN PROGRESS [J].
PAVCNIK, D ;
WRIGHT, KC ;
WALLACE, S .
RADIOLOGY, 1992, 183 (01) :151-154
[10]   Reduction of calcification by various treatments in cardiac valves [J].
Rao, KP ;
Shanthi, C .
JOURNAL OF BIOMATERIALS APPLICATIONS, 1999, 13 (03) :238-268