State of the art and challenges of time-of-flight PET

被引:222
作者
Conti, Maurizio [1 ]
机构
[1] Siemens Healthcare, Mol Imaging, Knoxville, TN 37932 USA
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2009年 / 25卷 / 01期
关键词
Positron emission tomography; Time of flight; TOF reconstruction; TOF PET; Scintillation detectors; TOF gain; Signal-to-noise ratio; POSITRON-EMISSION-TOMOGRAPHY; GAMMA-RAY SPECTROSCOPY; SCINTILLATION PROPERTIES; ENERGY-RESOLUTION; IMAGE-RECONSTRUCTION; TIMING RESOLUTION; TOF PET; 3-D PET; LSO; LUMINESCENCE;
D O I
10.1016/j.ejmp.2008.10.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
After a brief review of the history of time-of-flight (TOF) positron emission tomography (PET) instrumentation from the 1980s to present, the principles of TOF PET are introduced, the concept of time resolution and its effect on TOF gain in signal-to-noise ratio (SNR) are discussed. The factors influencing the time resolution of a TOF PET scanner are presented, with focus on the intrinsic properties of scintillators of particular interest for TOF PET. Finally, some open issues, challenges and achievements of today's TOF PET reconstruction are reviewed: the structure of the data organization, the choice of analytical. or iterative method, the recent experimental assessment of TOF image quality, and the most promising applications of TOF PET. (C) 2008 Associazione Itatiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1 / 11
页数:11
相关论文
共 83 条
[1]
ALLEMAND R, 1980, J NUCL MED, V21, P153
[2]
AYKAC M, 2006, IEEE T NUCT SCI, V53, P1184
[3]
A TECHNIQUE FOR THE CORRECTION OF SCATTERED RADIATION IN A PET SYSTEM USING TIME-OF-FLIGHT INFORMATION [J].
BENDRIEM, B ;
SOUSSALINE, F ;
CAMPAGNOLO, R ;
VERREY, B ;
WAJNBERG, P ;
SYROTA, A .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1986, 10 (02) :287-295
[4]
Optical spectroscopy and luminescence quenching of LuI3:Ce3+ [J].
Birowosuto, MD ;
Dorenbos, P ;
de Haas, JTM ;
van Eijk, CWE ;
Krämer, KW ;
Güdel, HU .
JOURNAL OF LUMINESCENCE, 2006, 118 (02) :308-316
[5]
Scintillation properties of LuI3:Ce3+-high light yield scintillators [J].
Birowosuto, MD ;
Dorenbos, P ;
van Eijk, CWE ;
Krämer, KW ;
Güdel, HU .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (04) :1114-1118
[6]
INSTRUMENTATION TRENDS IN NUCLEAR-MEDICINE [J].
BUDINGER, TF .
SEMINARS IN NUCLEAR MEDICINE, 1977, 7 (04) :285-297
[7]
BUDINGER TF, 1983, J NUCL MED, V24, P73
[8]
First experimental results of time-of-flight reconstruction on an LSO PET scanner [J].
Conti, M ;
Bendriem, B ;
Casey, M ;
Chen, M ;
Kehren, F ;
Michel, C ;
Panin, V .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19) :4507-4526
[9]
CONTI M, 2004, 2004 IEEE NUCL SCI S
[10]
CONTI M, 2009, 2008 S RAD MEAS APPL