Efficient protein digestion with peptide separation in a micro-device interfaced to electrospray mass spectrometry

被引:12
作者
Al-Lawati, H [1 ]
Watts, P [1 ]
Welham, KJ [1 ]
机构
[1] Univ Hull, Dept Chem, Kingston Upon Hull HU6 7RX, N Humberside, England
关键词
D O I
10.1039/b600752j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly efficient protein digestion device has been fabricated using commercially available immobilized trypsin on agarose beads, packed into a silica capillary and connected either directly to an electrospray mass spectrometer via a 'microtight T' connector, from which aqueous acetic acid ( 0.2%) was pumped, or via a monolithic column connected to the mass spectrometer ion source. Six proteins with molecular mass ranging from 2848 to 77703 Da were digested completely using this system. In the second set of experiments a short monolithic separation column was placed after the immobilized trypsin capillary and partial separation of the generated peptides was obtained. The detection limits were increased from the mu mol to pmol range by utilization of this separation column. Gradient elution, using a binary HPLC pump and a flow splitter, was used to optimize the peptide separation. This provided significantly enhanced resolution of the tryptic peptides but increased the analysis time to 30 minutes.
引用
收藏
页码:656 / 663
页数:8
相关论文
共 14 条
[1]   Reproducibility of a solid-phase trypsin microreactor for peptide mapping by capillary electrophoresis [J].
Bonneil, E ;
Mercier, M ;
Waldron, KC .
ANALYTICA CHIMICA ACTA, 2000, 404 (01) :29-45
[2]   Development of a bioreactor based on trypsin immobilized on monolithic support for the on-line digestion and identification of proteins [J].
Calleri, E ;
Temporini, C ;
Perani, E ;
Stella, C ;
Rudaz, S ;
Lubda, D ;
Mellerio, G ;
Veuthey, JL ;
Caccialanza, G ;
Massolini, G .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1045 (1-2) :99-109
[3]   Enhanced proteolytic activity enzymes in photopolymerized of covalently bound sol gel [J].
Dulay, MT ;
Baca, QJ ;
Zare, RN .
ANALYTICAL CHEMISTRY, 2005, 77 (14) :4604-4610
[4]  
Harrison DJ, 2003, LAB-ON-A CHIP: MINIATURIZED SYSTEMS FOR ( BIO ) CHEMICAL ANALYSIS AND SYNTHESIS, P249, DOI 10.1016/B978-044451100-3/50013-9
[5]   HYDROGEN-DEUTERIUM EXCHANGE ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY - A METHOD FOR PROBING PROTEIN CONFORMATIONAL-CHANGES IN SOLUTION [J].
KATTA, V ;
CHAIT, BT .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1993, 115 (14) :6317-6321
[6]   On-chip proteolytic digestion and analysis using "wrong-way-round" electrospray time-of-flight mass spectrometry [J].
Lazar, IM ;
Ramsey, RS ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2001, 73 (08) :1733-1739
[7]   ELECTROSPRAY MASS-SPECTROMETRIC STUDY OF MELITTIN TRYPSINOLYSIS BY A KINETIC APPROACH [J].
MIRGORODSKAYA, EP ;
MIRGORODSKAYA, OA ;
DOBRETSOV, SV ;
SHEVCHENKO, AA ;
DODONOV, AF ;
KOZLOVSKIY, VI ;
RAZNIKOV, VV .
ANALYTICAL CHEMISTRY, 1995, 67 (17) :2864-2869
[8]   A microscale electrospray interface incorporating a monolithic, poly(styrene-divinylbenzene) support for on-line liquid chromatography tandem mass spectrometry analysis of peptides and proteins [J].
Moore, RE ;
Licklider, L ;
Schumann, D ;
Lee, TD .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4879-4884
[9]   Analytical characterization of a facile porous polymer monolithic trypsin microreactor enabling peptide mass mapping using mass spectrometry [J].
Palm, AK ;
Novotny, MV .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (12) :1374-1382
[10]   Enzymatic microreactor-on-a-chip:: Protein mapping using trypsin immobilized on porous polymer monoliths molded in channels of microfluidic devices [J].
Peterson, DS ;
Rohr, T ;
Svec, F ;
Fréchet, JMJ .
ANALYTICAL CHEMISTRY, 2002, 74 (16) :4081-4088