A fast rebinning algorithm for 3D positron emission tomography using John's equation

被引:64
作者
Defrise, M [1 ]
Liu, XA [1 ]
机构
[1] Free Univ Brussels, Brussels, Belgium
关键词
D O I
10.1088/0266-5611/15/4/314
中图分类号
O29 [应用数学];
学科分类号
070104 ;
摘要
Volume imaging in positron emission tomography (PET) requires the inversion of the three-dimensional (3D) x-ray transform. The usual solution to this problem is based on 3D filtered-backprojection (FBP), but is slow. Alternative methods have been proposed which factor the 3D data into independent 2D data sets corresponding to the 2D Radon transforms of a stack of parallel slices. Each slice is then reconstructed using 2D FBP. These so-called rebinning methods are numerically efficient but are approximate. In this paper a new exact rebinning method is derived by exploiting the fact that the 3D x-ray transform of a function is the solution to the second-order partial differential equation first studied by John. The method is proposed for two sampling schemes, one corresponding to a pair of infinite plane detectors and another one corresponding to a cylindrical multi-ring PET scanner. The new FORE-J algorithm has been implemented for this latter geometry and was compared with the approximate Fourier rebinning algorithm FORE and with another exact rebinning algorithm, FOREX. Results with simulated data demonstrate a significant improvement in accuracy compared to FORE, while the reconstruction time is doubled. Compared to FOREX, the FORE-J algorithm is slightly less accurate but more than three times faster.
引用
收藏
页码:1047 / 1065
页数:19
相关论文
共 30 条
[1]  
Bailey DL, 1998, THEORY PRACTICE 3D P
[2]   3D PET USING A CONVENTIONAL MULTISLICE TOMOGRAPH WITHOUT SEPTA [J].
CHERRY, SR ;
DAHLBOM, M ;
HOFFMAN, EJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (04) :655-668
[3]   Fast reconstruction of 3D PET data with accurate statistical modeling [J].
Comtat, C ;
Kinahan, PE ;
Defrise, M ;
Michel, C ;
Townsend, DW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1083-1089
[4]  
DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717
[5]   Exact and approximate rebinning algorithms for 3-D PET data [J].
Defrise, M ;
Kinahan, PE ;
Townsend, DW ;
Michel, C ;
Sibomana, M ;
Newport, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :145-158
[6]   A FACTORIZATION METHOD FOR THE 3D X-RAY TRANSFORM [J].
DEFRISE, M .
INVERSE PROBLEMS, 1995, 11 (05) :983-994
[7]  
Edholm P. R., 1986, Proceedings of the SPIE - The International Society for Optical Engineering, V671, P8, DOI 10.1117/12.966672
[8]   LINOGRAMS IN IMAGE-RECONSTRUCTION FROM PROJECTIONS [J].
EDHOLM, PR ;
HERMAN, GT .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1987, 6 (04) :301-307
[9]   Incremental beamwise backprojection using geometrical symmetries for 3D PET reconstruction in a cylindrical scanner geometry [J].
Egger, ML ;
Joseph, C ;
Morel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :3009-3024
[10]   3D RECONSTRUCTION FOR A MULTIRING PET SCANNER BY SINGLE-SLICE REBINNING AND AXIAL DECONVOLUTION [J].
ERLANDSSON, K ;
ESSER, PD ;
STRAND, SE ;
VANHEERTUM, RL .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03) :619-629