Motion artifacts in subsecond conventional CT and electron-beam CT: Pictorial demonstration of temporal resolution

被引:29
作者
McCollough, CH [1 ]
Bruesewitz, MR [1 ]
Daly, TR [1 ]
Zink, FE [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
关键词
computed tomography; artifact; electron beam; image quality; phantoms; test objects;
D O I
10.1148/radiographics.20.6.g00nv131675
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To visually demonstrate the effective temporal resolution of subsecond conventional (slip-ring) and electron-beam computed tomographic (CT) systems, two phantoms containing high-contrast test objects were scanned with a slip-ring CT system (effective exposure time, 0.5 second) and an electron-beam CT system (exposure time, 0.1 second). Images were acquired of each phantom at rest, during translation along the x axis at speeds of 10-100 mm/sec, and during rotation about isocenter at speeds of 0.1 and 0.5 revolution per second. Motion artifacts and loss of spatial resolution were judged to be absent, noticeable, or severe. For 0.5-second conventional CT images, motion artifacts and loss of spatial resolution were noticeable at 10 mm/sec and 0.1 revolution per second and were severe at speeds greater than or equal to 20 mm/sec and at 0.5 revolution per second. For 0.1-second electron-beam CT scans, noticeable, but not severe, motion artifacts and loss of spatial resolution occurred at speeds between 40 and 100 mm/sec and at 0.5 revolution per second. Over the range of physiologic speeds examined, the images provide visually compelling evidence of the effect of improving temporal resolution in CT.
引用
收藏
页码:1675 / 1681
页数:7
相关论文
共 6 条
[1]   In-plane coronary arterial motion velocity: Measurement with electron-beam CT [J].
Achenbach, S ;
Ropers, D ;
Holle, J ;
Muschiol, G ;
Daniel, WG ;
Moshage, W .
RADIOLOGY, 2000, 216 (02) :457-463
[2]   Quantification of in-plane motion of the coronary arteries during the cardiac cycle: Implications for acquisition window duration for MR flow quantification [J].
Hofman, MBM ;
Wickline, SA ;
Lorenz, CH .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (03) :568-576
[3]   Coronary artery motion in electron beam tomography [J].
Mao, SS ;
Lu, B ;
Oudiz, RJ ;
Bakhsheshi, H ;
Liu, SCK ;
Budoff, MJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2000, 24 (02) :253-258
[4]   Quantitative pulmonary imaging: Spatial and temporal considerations in high-resolution CT [J].
Reinhardt, JM ;
Hoffman, EA .
ACADEMIC RADIOLOGY, 1998, 5 (08) :539-546
[5]   MINIMUM SCAN SPEEDS FOR SUPPRESSION OF MOTION ARTIFACTS IN CT [J].
RITCHIE, CJ ;
GODWIN, JD ;
CRAWFORD, CR ;
STANFORD, W ;
ANNO, H ;
KIM, Y .
RADIOLOGY, 1992, 185 (01) :37-42
[6]   Segmental and subsegmental pulmonary arteries: Evaluation with electron-beam versus spiral CT [J].
Schoepf, UJ ;
Helmberger, T ;
Holzknecht, N ;
Kang, DS ;
Bruening, RD ;
Aydemir, S ;
Becker, CR ;
Muehling, O ;
Knez, A ;
Haberl, R ;
Reiser, MF .
RADIOLOGY, 2000, 214 (02) :433-439