Closed-loop control of anesthesia using bispectral index - Performance assessment in patients undergoing major orthopedic surgey under combined general and regional anesthesia

被引:142
作者
Absalom, AR [1 ]
Sutcliffe, N [1 ]
Kenny, GN [1 ]
机构
[1] Glasgow Royal Infirm, Dept Anaesthesia, Glasgow, Lanark, Scotland
关键词
D O I
10.1097/00000542-200201000-00017
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: The Bispectral Index (BIS) is an electroencephalogram-derived measure of anesthetic depth. A closed-loop anesthesia system was built using BIS as the control variable, a proportional-integral-differential control algorithm, and a propofol target-controlled infusion system as (lie control actuator. Closed-loop performance was assessed in 10 adult patients. Methods: Ten adult patients scheduled to undergo elective hip or knee surgery were enrolled. An epidural cannula was inserted, and 0.5% bupivacaine was used to provide anesthesia to T8 before general anesthesia was Induced using (lie propofol target-controlled infusion system under manual control. After the start of surgery, when anesthesia was clinically adequate, automatic control of anesthesia was commenced using the BIS as the control variable. Adequacy of anesthesia during closed-loop control was assessed clinically and by calculating the median performance error, the median absolute performance error, and the mean offset of the control variable. Results: The median performance error and the median absolute performance error were 2.2 and 8.0%, respectively. Mean offset of the BIS from the set point was 0.9. Cardiovascular parameters were stable during closed-loop control. operating conditions were adequate in all patients but one, who began moving after 45 min of stable anesthesia. No patients reported awareness or recall of Intraoperative events. hi three patients, there was oscillation of the measured BIS around the set point. Conclusions: The system was able to provide clinically adequate anesthesia in 9 of 10 patients. Further studies are required to determine whether control performance can be improved by alterations to the gain factors or by using an effect site-targeted, target-controlled infusion propofol system.
引用
收藏
页码:67 / 73
页数:7
相关论文
共 22 条
  • [1] A comparison of spectral edge, delta power, and bispectral index as EEG measures of alfentanil, propofol, and midazolam drug effect
    Billard, V
    Gambus, PL
    Chamoun, N
    Stanski, DR
    Shafer, SL
    [J]. CLINICAL PHARMACOLOGY & THERAPEUTICS, 1997, 61 (01) : 45 - 58
  • [2] Prediction of movement at laryngeal mask airway insertion: comparison of auditory evoked potential index, bispectral index, spectral edge frequency and median frequency
    Doi, M
    Gajraj, RJ
    Mantzaridis, H
    Kenny, GNC
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1999, 82 (02) : 203 - 207
  • [3] Awareness during anesthesia
    Ghoneim, MM
    [J]. ANESTHESIOLOGY, 2000, 92 (02) : 597 - 602
  • [4] Bispectral analysis measures sedation and memory effects of propofol, midazolam, isoflurane, and alfentanil in healthy volunteers
    Glass, PS
    Bloom, M
    Kearse, L
    Rosow, C
    Sebel, P
    Manberg, P
    [J]. ANESTHESIOLOGY, 1997, 86 (04) : 836 - 847
  • [5] JACOBS JR, 1993, ANESTH ANALG, V76, P1
  • [6] Performance assessment of an adaptive model-based feedback controller: Comparison between atracurium, mivacurium, rocuronium and vecuronium
    Kansanaho, M
    Olkkola, KT
    [J]. INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1996, 13 (04): : 217 - 224
  • [7] Electroencephalographic derivatives as a tool for predicting the depth of sedation and anesthesia induced by sevoflurane
    Katoh, T
    Suzuki, A
    Ikeda, K
    [J]. ANESTHESIOLOGY, 1998, 88 (03) : 642 - 650
  • [8] Closed-loop control of propofol anaesthesia
    Kenny, GNC
    Mantzaridis, H
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (02) : 223 - 228
  • [9] Middle latency auditory evoked responses and electroencephalographic derived variables do not predict movement to noxious stimulation during 1 minimum alveolar anesthetic concentration isoflurane nitrous oxide anesthesia
    Kochs, E
    Kalkman, CJ
    Thornton, C
    Newton, D
    Bischoff, P
    Kuppe, H
    Abke, J
    Konecny, E
    Nahm, W
    Stockmanns, G
    [J]. ANESTHESIA AND ANALGESIA, 1999, 88 (06) : 1412 - 1417
  • [10] Kurita T, 1999, ANESTHESIOLOGY, V91, pU258