Comparative study of five instruments measuring stratum corneum hydration (Corneometer CM 820 and CM 825, Skicon 200, Nova DPM 9003, DermaLab). Part II. In vivo

被引:69
作者
Fluhr, Joachim W. [1 ]
Gloor, Max [1 ]
Lazzerini, Sabrina [2 ]
Kleesz, Peter [3 ]
Grieshaber, Romano [3 ]
Berardesca, Enzo [2 ]
机构
[1] Stadt Klinikum Karlsruhe, Dept Dermatol, Moltkestr 120, D-76133 Karlsruhe, Germany
[2] Univ Pavia, ZRCCS, Policlin S Matteo, Dept Dermatol, Pavia, Italy
[3] BG Nahrungsmittel & Gaststatten, Res Ctr, Mannheim, Germany
关键词
bioengineering; capacitance; conductance; stratum corneum hydration; water holding capacity;
D O I
10.1111/j.1600-0846.1999.tb00127.x
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background/Aims: The hydration state of stratum corneum can be measured with different instruments. In the present study, five instruments were compared under different experimental in vivo conditions: Corneometer CM 820 and CM 825 (both capacitance based), Skicon 200 (conductance), Nova DPM (impedance based capacitance) and DermaLab (capacitance). Methods: Seven anatomical sites were tested, a sorption-desorption test (SDT) and a moisture accumulation test (MAT) were performed. Different hydration states were created on the volar forearms by occluded application of an oil-in-water (o/w) emulsion, glycerol and urea containing oil-in-water (o/w) emulsion, an untreated control site and a site after lipid extraction with acetone. Measurements were performed 15 min after removal of occlusion. Results: The coefficient of variation for CM 820 ranged between 9.3% and 23.7%; for CM 825 between 14.9% and 36.8%; for Skicon 200 between 53.0% and 167.3%; for Nova DPM between 36.4% and 95.8%; for DermaLab between 23.6% and 45.6%. All devices showed a highly significant correlation (P<0.0001) with a high Spearman coefficient r ranging between 0.7928 and 0.9358. Conclusions: CM 820 and CM 825 seem to differentiate more precisely than Nova DPM, Skicon 200 and DermaLab in dry conditions according to previous findings, i.e., that conductance measurements might be more sensitive for hydrated skin and capacitance measurements might be more sensitive for dry skin conditions. The CM 825, the Skicon 200 and the Nova DPM are suitable for dynamic measurements of hydration parameters of the stratum corneum.
引用
收藏
页码:171 / 178
页数:8
相关论文
共 24 条
[11]  
HASHIMOTOKUMASAKA K, 1993, ACTA DERM-VENEREOL, V73, P335
[12]   INVITRO COMPARISON OF WATER-HOLDING CAPACITY OF THE SUPERFICIAL AND DEEPER LAYERS OF THE STRATUM-CORNEUM [J].
HASHIMOTOKUMASAKA, K ;
HORII, I ;
TAGAMI, H .
ARCHIVES OF DERMATOLOGICAL RESEARCH, 1991, 283 (05) :342-346
[13]   FRICTION, CAPACITANCE AND TRANSEPIDERMAL WATER-LOSS (TEWL) IN DRY ATOPIC AND NORMAL SKIN [J].
LODEN, M ;
OLSSON, H ;
AXELL, T ;
LINDE, YW .
BRITISH JOURNAL OF DERMATOLOGY, 1992, 126 (02) :137-141
[14]   ELECTRICAL DETERMINATION OF WATER-CONTENT AND CONCENTRATION PROFILE IN A SIMULATION-MODEL OF INVIVO STRATUM-CORNEUM [J].
OBATA, M ;
TAGAMI, H .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1989, 92 (06) :854-859
[15]   METHOD TO EVALUATE SKIN MOISTURIZERS INVIVO [J].
RIETSCHEL, RL .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1978, 70 (03) :152-155
[16]  
Rogiers V., 1990, COSMETICS TOILETRIES, V105, P73
[17]   WATER SORPTION-DESORPTION TEST OF THE SKIN INVIVO FOR FUNCTIONAL ASSESSMENT OF THE STRATUM-CORNEUM [J].
TAGAMI, H ;
KANAMARU, Y ;
INOUE, K ;
SUEHISA, S ;
INOUE, F ;
IWATSUKI, K ;
YOSHIKUNI, K ;
YAMADA, M .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1982, 78 (05) :425-428
[18]   STRATUM-CORNEUM DYNAMIC FUNCTION MEASUREMENTS AFTER MOISTURIZER OR IRRITANT APPLICATION [J].
TREFFEL, P ;
GABARD, B .
ARCHIVES OF DERMATOLOGICAL RESEARCH, 1995, 287 (05) :474-479
[19]   COMPARATIVE INVESTIGATIONS ON THE WATER-CONTENT OF THE STRATUM-CORNEUM USING DIFFERENT METHODS OF MEASUREMENT [J].
TRIEBSKORN, A ;
GLOOR, M ;
GREINER, F .
DERMATOLOGICA, 1983, 167 (02) :64-69
[20]  
Van Neste D, 1991, J Dermatol Sci, V2, P119, DOI 10.1016/0923-1811(91)90021-O