Inversion recovery TrueFISP:: Quantification of T1, T2, and spin density

被引:201
作者
Schmitt, P
Griswold, MA
Jakob, PM
Kotas, M
Gulani, V
Flentje, M
Haase, A
机构
[1] Univ Wurzburg, Inst Phys, Klin & Poliklin Strahlentherapie, D-97074 Wurzburg, Germany
[2] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
关键词
T-1; quantification; T-2; measurement; inversion recovery; TrueFISP; steady state free precession; SSFP;
D O I
10.1002/mrm.20058
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel procedure is proposed to extract T-1, T-2, and relative spin density from the signal time course sampled with a series of TrueFISP images after spin inversion. Generally, the recovery of the magnetization during continuous TrueFISP imaging can be described in good approximation by a three parameter monoexponential function S(t) = S-stst(1-INV exp(-t/T-1(*)). This apparent relaxation time T-1(*) less than or equal to T-1 depends on the flip angle as well as on both T-1 and T-2. Here, it is shown that the ratio T-1/T-2 can be directly extracted from the inversion factor INV, which describes the relation of the signal value extrapolated to t = 0 and the steady-state signal. Analytical expressions are given for the derivation of T-1, T-2, and relative spin density directly from the fit parameters. Phantom results show excellent agreement with single point reference measurements. In human volunteers T-1, T-2, and spin density maps in agreement with literature values were obtained. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:661 / 667
页数:7
相关论文
共 24 条
[1]   T1 AND T2 MEASUREMENTS ON A 1.5-T COMMERCIAL MR IMAGER [J].
BREGER, RK ;
RIMM, AA ;
FISCHER, ME ;
PAPKE, RA ;
HAUGHTON, VM .
RADIOLOGY, 1989, 171 (01) :273-276
[2]   EVALUATION OF PROTON DENSITY BY MAGNETIC-RESONANCE-IMAGING - PHANTOM EXPERIMENTS AND ANALYSIS OF MULTIPLE COMPONENT PROTON TRANSVERSE RELAXATION [J].
BRIX, G ;
SCHAD, LR ;
LORENZ, WJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (01) :53-66
[3]   THE DIFFUSION SENSITIVITY OF FAST STEADY-STATE FREE PRECESSION IMAGING [J].
BUXTON, RB .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (02) :235-243
[4]   Cine MR angiography of the heart with segmented true fast imaging with steady-state precession [J].
Carr, JC ;
Simonetti, O ;
Bundy, J ;
Li, DB ;
Pereles, S ;
Finn, JP .
RADIOLOGY, 2001, 219 (03) :828-834
[5]  
DEIMLING M, 1994, P 2 ANN M ISMRM
[6]   PHASE AND INTENSITY ANOMALIES IN FOURIER TRANSFORM NMR [J].
FREEMAN, R ;
HILL, HDW .
JOURNAL OF MAGNETIC RESONANCE, 1971, 4 (03) :366-&
[7]   STEADY-STATE FREE PRECESSION IMAGING IN THE PRESENCE OF MOTION - APPLICATION FOR IMPROVED VISUALIZATION OF THE CEREBROSPINAL-FLUID [J].
HAACKE, EM ;
WIELOPOLSKI, PA ;
TKACH, JA ;
MODIC, MT .
RADIOLOGY, 1990, 175 (02) :545-552
[8]  
Haacke EM, 1999, MAGNETIC RESONANCE I
[9]   Characterization and reduction of the transient response in steady-state MR imaging [J].
Hargreaves, BA ;
Vasanawala, SS ;
Pauly, JM ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (01) :149-158
[10]   A fast 3D Look-Locker method for volumetric T1 mapping [J].
Henderson, E ;
McKinnon, G ;
Lee, TY ;
Rutt, BK .
MAGNETIC RESONANCE IMAGING, 1999, 17 (08) :1163-1171