Shell and small particles; Evaluation of new column technology

被引:61
作者
Fekete, Szabolcs [1 ]
Fekete, Jeno [2 ]
Ganzler, Katalin [1 ]
机构
[1] Gedeon Richter Plc, Formulat Dev, Budapest, Hungary
[2] Budapest Univ Technol & Econ, Dept Inorgan & Analyt Chem, H-1111 Budapest, Hungary
关键词
Column efficiency; Shell particles; Peak capacity; Kinetic plot; Ascentis Express; PERFORMANCE LIQUID-CHROMATOGRAPHY; PEAK-CAPACITY; SILICA PARTICLES; MASS-TRANSFER; POROSITY; KINETICS; POLYMER; HPLC;
D O I
10.1016/j.jpba.2008.10.009
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of 5 cm long columns packed with shell particles was compared to totally porous sub-2 mu m particles in gradient and isocratic elution separations of hormones (dienogest, finasteride, gestodene, levonorgestrel, estradiol, ethinylestradiol, noretistherone acetate, bicalutamide and tibolone). Peak capacities around 140-150 could be achieved in 25 min with the 5 cm long columns. The Ascentis Express column (packed with 2.7 mu m shell particles) showed similar efficiency to sub-2 mu m particles under gradient conditions. Applying isocratic separation, the column of 2.7 pm shell particles had a reduced plate height minimum of approximately h = 1.6. It was much smaller than obtained with totally porous particles (h approximate to 2.8). The impedance time also proved more favorable with 2.7 mu m shell particles than with totally porous particles. The influence of extra-column volume on column efficiency was investigated. The extra-column dispersion of the chromatographic system may cause a shift of the HETP curves. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:64 / 71
页数:8
相关论文
共 30 条
[1]   A CHROMATOGRAPHIC STUDY OF THE LEWIS ACID-BASE CHEMISTRY OF ZIRCONIA SURFACES [J].
BLACKWELL, JA ;
CARR, PW .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1991, 14 (15) :2875-2889
[2]   STANDARDIZATION OF TEST CONDITIONS FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COLUMNS [J].
BRISTOW, PA ;
KNOX, JH .
CHROMATOGRAPHIA, 1977, 10 (06) :279-289
[3]  
Cazes J., 2002, CHROMATOGRAPHY THEOR
[4]  
Claessens H A., 2004, Reduction of Analysis Times in HPLC at Elevated Column Temperatures
[5]  
Desmet G, 2008, LC GC EUR, V21, P310
[6]   Future of high pressure liquid chromatography: Do we need porosity or do we need pressure? [J].
Desmet, Gert ;
Cabooter, Deirdre ;
Gzil, Piotr ;
Verelst, Harry ;
Mangelings, Debby ;
Vander Heyden, Yvan ;
Clicq, David .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1130 (01) :158-166
[7]   Characteristics of superficially-porous silica particles for fast HPLC:: Some performance comparisons with sub-2-μm particles [J].
DeStefano, J. J. ;
Langlois, T. J. ;
Kirkland, J. J. .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (03) :254-260
[8]  
FELINGER A, 1998, SIGNAL PROCESSING CH
[9]   Comparative study of the performance of columns packed with several new fine silica particles Would the external roughness of the particles affect column properties? [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1166 (1-2) :30-46
[10]   Comparison between the efficiencies of columns packed with fully and partially porous C18-bonded silica materials [J].
Gritti, Fabrice ;
Cavazzini, Alberto ;
Marchetti, Nicola ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1157 (1-2) :289-303