Effective dose in paediatric computed tomography

被引:64
作者
Chapple, CL [1 ]
Willis, S [1 ]
Frame, J [1 ]
机构
[1] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
关键词
D O I
10.1088/0031-9155/47/1/308
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is limited data currently available for making dose and risk assessments for paediatric patients undergoing computed tomographic examination. A method has been developed to correlate the risk-related quantity, effective dose, to the more simply derived quantity dose-length product. This involved scanning a series of paediatric anthropomorphic phantoms containing thermoluminescent dosimeters to measure effective dose for scans of various anatomic regions. The quantity effective dose per dose-length product was calculated and plotted as a function of patient size. This showed a simple exponential relationship. and equations of fit were derived to enable the calculation of effective dose for a patient of any size. Measurements carried out on a second scanner and for alternative scan volumes indicated that the method could be generally utilized.
引用
收藏
页码:107 / 115
页数:9
相关论文
共 19 条