Determination of cadmium, mercury and lead in seawater by electrothermal vaporization isotope dilution inductively coupled plasma mass spectrometry

被引:107
作者
Liu, HW [1 ]
Jiang, SJ [1 ]
Liu, SH [1 ]
机构
[1] Natl Sun Yat Sen Univ, Dept Chem, Kaohsiung 804, Taiwan
关键词
electrothermal vaporization; isotope dilution; inductively coupled plasma-mass spectrometry (ICP-MS); Cd; Hg; Pb; sea water;
D O I
10.1016/S0584-8547(99)00081-6
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Electrothermal vaporization isotope dilution inductively coupled plasma mass spectrometry (ETV-ID-ICP-MS) has been applied to the determination of Cd, Hg and Pb in seawater samples. The isotope ratios of the elements studied in each analytical run were calculated from the peak areas of each isotope. Various modifiers were tested for the best signal of these elements. After preliminary studies, 0.15% m/v TAC and 4% v/v HCl were added to the sample solution to work as the modifier. The ETV-ID-ICP-MS method has been applied to the determination of Cd, Hg and Pb in NASS-4 and CASS-3 reference seawater samples and seawater samples collected from Kaohsiung area. The results for reference sample NASS-4 and CASS-3 agreed satisfactorily with the reference values. Results for other samples determined by isotope dilution and method of standard additions agreed satisfactorily. Detection limits were approximately 0.002, 0.005 and 0.001 ng ml(-1) for Cd, Hg and Pb in seawater, respectively, with the ETV-ICP-MS method. Precision between sample replicates was better than 20% for most of the determinations. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1367 / 1375
页数:9
相关论文
共 30 条
[21]   Determination of copper, cadmium and lead in sediment samples by slurry sampling electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Liaw, MJ ;
Jiang, SJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (08) :555-560
[22]   Determination of mercury in fish samples by slurry sampling electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Liaw, MJ ;
Jiang, SJ ;
Li, YC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (06) :779-785
[23]  
LIAW MJ, 1909, THESIS NATL SUN YAT
[24]   APPLICATION OF ISOTOPE-DILUTION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY TO THE ANALYSIS OF MARINE-SEDIMENTS [J].
MCLAREN, JW ;
BEAUCHEMIN, D ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1987, 59 (04) :610-613
[25]   A LOW CONTAMINATION CHELEX-100 TECHNIQUE FOR SHIPBOARD PRECONCENTRATION OF HEAVY-METALS IN SEAWATER [J].
PAI, SC ;
FANG, TH ;
CHEN, CTA ;
JENG, KL .
MARINE CHEMISTRY, 1990, 29 (04) :295-306
[26]   Direct determination of trace metals in sea-water by inductively coupled plasma mass spectrometry [J].
Rosland, E ;
Lund, W .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (11) :1239-1244
[27]   PARTICLE-SIZE DISTRIBUTION OF SAMPLE TRANSPORTED FROM AN ELECTROTHERMAL VAPORIZER TO AN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETER [J].
SPARKS, CM ;
HOLCOMBE, J ;
PINKSTON, TL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (13) :1607-1615
[28]   Liquid-liquid extraction of cadmium by sodium diethyldithiocarbamate from biological matrixes for isotope dilution inductively coupled plasma mass spectrometry [J].
Vanderpool, RA ;
Buckley, WT .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :652-659
[29]  
WEI MT, UNPUB DETERMINATION
[30]   Determination of inorganic and total mercury in biological tissues by electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Willie, SN ;
Gregoire, DC ;
Sturgeon, RE .
ANALYST, 1997, 122 (08) :751-754