Assessment of changes in dual-energy X-ray absorptiometry performance following a system upgrade

被引:2
作者
Simmons, A [1 ]
Barrington, SF [1 ]
Archbold, LJ [1 ]
ODoherty, MJ [1 ]
Coakley, AJ [1 ]
机构
[1] KENT & CANTERBURY HOSP,E KENT OSTEOPOROSIS SCREENING & RES UNIT,CANTERBURY CT1 3NG,KENT,ENGLAND
关键词
D O I
10.1097/00006231-199604000-00161
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
System upgrades are a constant feature of many highly technical specialities and are normally assumed to have a negligible effect on patient data. We report on the assessment of an upgrade made to a Norland XR-26 dual-energy X-ray absorptiometry (DXA) system. The upgraded version (2.3) consisted of new hip positioning aids, hip image analysis software and updates to the normal ranges. Measurements of femoral neck and trochanteric bone mineral density correlated well between versions (r = 0.969 and r = 0.971), those for femoral neck and trochanteric bone mineral content less well (r = 0.956 and r = 0.911), and Ward's triangle values and femoral neck area measurements were poorly correlated (r = 0.719-0.881). The upgrade was claimed to improve precision while reducing analysis and imaging time. Improved precision was not demonstrated, with good precision found both before and after the upgrade. Correlation between femoral neck Z-scores were good for the European normal range (r = 0.970), but unacceptable for the UK range (r = 0.769). These correlations were not close enough to allow patients scanned with the old version to be followed up with scans using the upgrade. The upgrade was, however, preferred for new attenders because of the improved speed and automation of hip analysis, without loss of precision. The most important implication in terms of patient management was the change made to the UK normal range, which had a very large effect on the number of patients deemed to be at risk from osteoporosis. It is essential that any upgrade introduced to an existing DXA system is carefully evaluated, since it may affect system performance and patient results.
引用
收藏
页码:331 / 341
页数:11
相关论文
共 28 条
[1]  
ARAI H, 1990, IMAGE TECHNOL INFORM, V22, P1
[2]   EFFECT OF SEASON ON PHYSICAL-ACTIVITY SCORE, BACK EXTENSOR MUSCLE STRENGTH, AND LUMBAR BONE-MINERAL DENSITY [J].
BERGSTRALH, EJ ;
SINAKI, M ;
OFFORD, KP ;
WAHNER, HW ;
MELTON, LJ .
JOURNAL OF BONE AND MINERAL RESEARCH, 1990, 5 (04) :371-377
[3]   DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS [J].
BLAKE, GM ;
PARKER, JC ;
BUXTON, FMA ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :902-906
[4]   DUAL ENERGY X-RAY ABSORPTIOMETRY MEASUREMENT OF BONE-MINERAL CONTENT IN NEWBORNS - VALIDATION OF THE TECHNIQUE [J].
BRAILLON, PM ;
SALLE, BL ;
BRUNET, J ;
GLORIEUX, FH ;
DELMAS, PD ;
MEUNIER, PJ .
PEDIATRIC RESEARCH, 1992, 32 (01) :77-80
[5]   SPINAL BONE-DENSITY MEASURED WITH A 5-S SCAN [J].
BUXTON, FMA ;
BLAKE, GM ;
PARKER, JC ;
LEWIS, MK ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (783) :275-277
[6]  
Duboeuf F, 1991, Osteoporos Int, V1, P242
[7]   FACTORS INFLUENCING SHORT-TERM PRECISION OF DUAL X-RAY BONE ABSORPTIOMETRY (DXA) OF SPINE AND FEMUR [J].
ENGELKE, K ;
GLUER, CC ;
GENANT, HK .
CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (01) :19-25
[8]   CROSS-CALIBRATION OF DXA EQUIPMENT - UPGRADING FROM A HOLOGIC QDR 1000/W TO A QDR-2000 [J].
FAULKNER, KG ;
GLUER, CC ;
ESTILO, M ;
GENANT, HK .
CALCIFIED TISSUE INTERNATIONAL, 1993, 52 (02) :79-84
[9]  
GENANT HK, 1994, J BONE MINER RES, V9, P1503
[10]   COMPARATIVE-ASSESSMENT OF DUAL-PHOTON ABSORPTIOMETRY AND DUAL-ENERGY RADIOGRAPHY [J].
GLUER, CC ;
STEIGER, P ;
SELVIDGE, R ;
ELLIESENKLIEFOTH, K ;
HAYASHI, C ;
GENANT, HK .
RADIOLOGY, 1990, 174 (01) :223-228