Nonlinear calibration of a laser stripe profiler

被引:36
作者
McIvor, AM [1 ]
机构
[1] Reveal Ltd, Auckland, New Zealand
关键词
laser profiler; calibration; nonlinear methods; least squares;
D O I
10.1117/1.1416694
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The calibration of a laser stripe profiler consisting of a laser stripe projector, camera, and linear motion table is considered. A nonlinear system model is used, which accommodates radial distortion in the camera lens. The way in which stripe data is extracted from the camera images leads to a natural formulation of the calibration problem as a nonlinear least squares problem. This can then be solved using standard techniques. The use of this nonlinear model reduces the error in the generated 3-D data by over an order of magnitude. (C) 2002 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页码:205 / 212
页数:8
相关论文
共 16 条
[1]  
BOLLES RC, 1981, 249 SRI STANF U
[2]   Photogrammetric calibration and accuracy evaluation of a cross-pattern stripe projector [J].
Brenner, C ;
Böhm, J ;
Gühring, J .
VIDEOMETRICS VI, 1998, 3641 :164-172
[3]  
Chen C. H., 1987, Proceedings of the 1987 IEEE International Conference on Robotics and Automation (Cat. No.87CH2413-3), P807
[4]  
Chiarella M., 1989, P SOC PHOTO-OPT INS, V1194, P176
[5]  
DEPIERO FW, 1995, ADV COMPUT, V43, P243
[6]  
Fisher RB, 1991, P BRIT MACH VIS C, P217
[7]   Triangulation [J].
Hartley, RI ;
Sturm, P .
COMPUTER VISION AND IMAGE UNDERSTANDING, 1997, 68 (02) :146-157
[8]  
JARVIS R, 1993, ADV IMAGE COMMUNICAT, P17
[9]  
MCIVOR AM, 1995, IMAGE VISION COMPUT, P167
[10]  
PAJDLA T, 1995, K335199598 CZECH TEC